M y L a b
–
S A F E T Y A N D S T A N D A R D
2-1
2 - Patient Safety
Electrical Safety
Warnings
The system must be properly grounded to prevent shock hazards.
Protection is provided by grounding the chassis with a three-wire cable
and plug; the system must also be powered through a properly
grounded receptacle.
Do not replace the system fuses with types different from those
specified by the
MyLab
“
Getting Started” manual.
Mobile configurations provide insulated plugs and connectors to
manage optional hard copy devices (VTR, printers). Follow the
instructions in the “Getting Started” manual to install such a device.
Incorrect connections may compromise the electrical safety of the
system.
If the Operator plans to use hard-copy devices with a portable model,
and one plans to utilize hard-copy devices, read and carefully follow
the instructions in the “Getting Started” manual to install such devices.
Incorrect connections or use of peripherals with improper safety
characteristics may compromise the electrical safety of the system.
MyLab
models are not watertight and provides a class IP(X)0 degree of
protection to liquids; do not expose the system to rain or moisture.
Avoid placing liquid containers on the system.
Remove probes and electrocardiography leads from patient contact
before applying a high voltage defibrillation pulse.
MyLab
systems use high frequency signals. Pacemakers could interfere
with these signals. The user should be aware of this minimal potential
hazard and immediately turn the unit off if interference with the
pacemaker operation is noted or suspected.
While using the system in combination with high frequency devices
(like electro-surgical units), be aware that a failure in the surgical device
or a damage to the transducer lens can cause electro-surgical currents
that can burn the patient. Thoroughly check the system and the probe
Chapter
2
W A R N I N G S
Observe the
following warnings
for maximum safety
G
G
S
S