Room Sensing Feature
The PK-CART ULTRA is equipped with Room Sensing
Technology in which the device will calculate the distance
between walls and calculate the required machine run-time
depending on Operation Mode (Bacteria or Spores) selected.
1. From Home Screen, press Right Arrow at bottom right of
screen.
2. Room Sensing Screen will appear with distance shown
between each measuring laser and wall. This screen is
for observation only.
a.
Each laser will measure up to a 16’ radius for a total of
32’ forward and backward and 32’ side to side
b.
The machine will always calculate run-time based on the
longest measured point ensuring a sufficient treatment
dose
c.
If any distance measure is greater than 16’, the machine
will automatically run 30-minute cycle on Bacteria
setting and 60-minute cycle on Spores setting.
3. Cycle time is displayed on Room Sensing screen and
can be observed automatically adjusting with movement
of the machine and selection of target Operation mode
4. Press “Home” button to return to Start Cycle screen for
initiating treatment
Cycle stop displays for normal system operation are covered
on this page.
Emergency Stop
Should cycle deactivation occur by engagement of the
Emergency Stop button, the following method will be required
to reset the system:
1. Observe Emergency Stop Active status on control screen
2. Release Emergency Stop by pulling up on the red knob
3. Select HOME button to return to main screen
4. Press START CYCLE to restart normal operation of system
5. Leave the area.
PIR Sensor Cycle Interruption
Should cycle interruption occur by means of a PIR
sensor activation, the following will be required:
1. Observe the interruption status type on the control panel
2. Ensure no one is in treatment area
3. Select HOME button to return to main screen
4. Press START CYCLE to restart normal safe operation of system
5. Leave the area.
OPERATION SETTINGS
CYCLE STOP NOTIFICATIONS