background image

 

Sana comfort 150 and 250 REHA 

 

 

Article no.: 2.520064 rev.: b 

Page 13 of 24 

 

6

 

Start-up and initial preparation 

6.1

 

Blood pressure recorder for 250 

In order to conduct exercise testing correctly, it is of utmost importance to measure physical perform-

ance data and data from the ECG measurement as well as simultaneously measuring and recording blood 

pressure data to determine the reaction of the circulatory system to increased exertion. 

For this purpose, ergosana has developed an extremely accurate blood pressure measuring system that 
is not susceptible to interference. It has been integrated into this ergometer and uses a so-called indirect 

method of blood pressure measurement. The Korotkoff sound, which is created by the air being forced 

out of the cuff as blood flows through the area of compression, is recorded along with several other im-

portant parameters of critical importance to attain accurate measurement. These measurements are 
evaluated in milliseconds by an internal digital evaluation system and shown in the ergometer's display 

as systole and diastole. The pulse rate is also determined during measurement and likewise shown in the 

display. At the same time as they are shown in the display, the measurements can also be transferred to 

a peripheral device such as an ECG or pulmonary function unit for evaluation and recording over an RS 
232 interface. 

The blood pressure cuff is the measurement sensor for blood pressure. Despite the perfectly functioning 

measuring system, it remains critically important that the cuff is placed on the arm correctly and care-

fully. 

According to international agreement, the blood pressure should be measured on the left arm, which is 

near the heart, as the flow impedance level is lowest there. An exception to this rule is formed by approx. 

1 to 2 per cent of test persons on whom the Korotkoff sound cannot be measured due to vascular phe-

nomena. The cuff is placed on the right arm of such patients. 

 

 

Please note that the cuff's air tube must be fixed in a way that prevents it from dashing against the 
ergometer. This is to prevent unnecessary artefacts that might affect the measurement's accuracy. 

6.2

 

Applying the cuff for 250 

The microphone is positioned so that it lies on the brachial artery, the largest arm artery. The location of 

the microphone in the cuff is marked with a red fabric tag. 

The ideal location for the microphone is approx. 2 centimetres above the elbow joint on the inside of the 
arm, below the biceps. The cuff must be put on so it is tight and cannot shift out of position during the 

movement created during the stress test. 

The cuff is inflated rapidly at the start of the measurement. The blood pressure and pulse rate are al-

ready roughly measured during pumping and the inflation pressure is determined. 

After the systolic pressure value has been attained, the air is released from the cuff at a rate of 3 mmHg 

per heartbeat. 

This procedure guarantees approximately equal measuring times despite the rising pulse rate during ex-

ertion. 

The blood pressure measurement should not exceed a maximum total length of 45 seconds. One minute 

is indicated as the shortest measurement interval. A measurement interval of 2 or 3 minutes is preferable 

in most cases. 

The blood pressure values are shown in the ergometer's graphic display together with the load and pulse 

rate graphs. 

Summary of Contents for Sana comfort 150 REHA

Page 1: ...Article no 2 520064 rev b Seat ergometer Sana comfort 150 250 REHA User Guide...

Page 2: ...en designed and manufactured in accordance with the relevant requirements of the EC directive 93 42 EEC This declaration loses its validity if the above devices are modified without ergosana s consent...

Page 3: ...notes 12 5 1 Precautions during operation 12 5 2 Safety precautions when operating with other devices 12 5 3 Maintenance safety precautions 12 5 4 Interference 12 6 Start up and initial preparation 13...

Page 4: ...pectoris serious hypertonia at rest carditis insufficiency of the heart serious valvular heart defect serious cardiac arrhythmia at rest aortic aneu rysm or other manifest cardiovascular diseases 1 3...

Page 5: ...essories Every device comes with Power cable with European plug Blood pressure cuff for 250 User guide Inspection report 2 3 Potential equalisation A standard potential equalisation stud is located on...

Page 6: ...250 Indirectly with a specific modified measuring system based on R R and computer analysis including maximal suppression of artefacts during ergometry Automatic deflation rate of 3 mmHg pulse Meas u...

Page 7: ...environment and the humidity is at least 30 especially if the floors are synthetic 5 If disorders should nevertheless occur especially in the vicinity of devices labelled with the symbol non ionic el...

Page 8: ...ust the device so that it is flush with the floor The ergometer is then fully stabilised 3 2 2 Seat adjustment Mechanical seat adjustment The seat is infinitely variable horizontally in order to accur...

Page 9: ...he complete range of information is located on the front side under a foil key board with a see through window The control elements for the adjustment and operation of the ergome ter are located on th...

Page 10: ...nt measurement data is shown in the display in alphanumeric and graphic forms It is thus especially easy to monitor exercise testing The cursor can be moved through the selection menu in the display w...

Page 11: ...ilable for larger arm diameters A microphone is installed in a microphone pocket on the inside of the cuff It serves the transmission of the blood pressure sound The connection cable with an air and a...

Page 12: ...clamps with the levers positioned vertically downward Secure clamping is en sured when the handles are then again turned to this position during the adjustment procedure after every movement The hold...

Page 13: ...care fully According to international agreement the blood pressure should be measured on the left arm which is near the heart as the flow impedance level is lowest there An exception to this rule is...

Page 14: ...metry is highlighted with a black bar This means the item is activated It is possible to select an exercise program directly from here and begin ergometry Beforehand however the five varying exercise...

Page 15: ...on ok and the window containing the settings menu for ergome try program no 1 appears Check the existing settings for Type of load Stage or Ramp Initial load Increase in load Stage time Blood pressure...

Page 16: ...ber Initial load watts Increase in load watts Stage time min BP interval min Recovery load watts 1 30 10 1 2 20 2 25 25 2 2 25 3 50 25 2 2 25 4 50 50 3 3 50 5 75 50 3 3 50 The program always returns t...

Page 17: ...the ergometer via an interface cable For our ergometers this is a type RS 232 interface which is isolated to ensure the patient s safety The appropriate baud rate must be selected in the program Sett...

Page 18: ...ing a training program on the ergometer To define a training program select the menu option Define program Then select the menu option Train ing using the arrow keys The below settings panel is displa...

Page 19: ...the training phase It is important that the body heart and circulation are trained but no overload is exerted The pulse steady state program is a very easy means to achieve this If the performance dec...

Page 20: ...used in combination with a number of rehabilitation systems To be able to implement this upgrade the ergometer has the following additional functions or aggregates a ECG amplifier incl USB converter...

Page 21: ...bus If several ergometers are in use the data transfer for the rehabilitation control program Sana Sprint remote operation is performed via the USB bus only Please note The RS 232 interface is used fo...

Page 22: ...th a soft cloth that is dry or moist Commercially available cleaning agents for household appliances can be used The saddle should be cleaned with a leatherette cleaner It is imperative to make sure t...

Page 23: ...tch After a continuity check change the fuses if necessary Return the fuse switch to the well and press it into the lock in position Replace fuses only with other fuses of the same type with the same...

Page 24: ...gents are trained in the service of our devices Please contact one of these specialist dealers if your unit requires servicing If this is not possible please contact the company s central service depa...

Page 25: ......

Reviews: