Information for the operator
10/80
2.6.1
Assembly
DANGER!
Personal danger due to electric shock or burns!
–
The assembly and electrical connection must comply
with the national regulations.
–
The on-site electrical installation must be fitted with an
easily accessible all-pole isolating device (master
switch), complying with overvoltage category III. This
means that each pole must have a contact opening
width complying with the conditions of overvoltage
category III for full isolation.
–
If the connection is established via a plug-in connection
then the connection system according to EN 60309
must be utilized.
–
Assembly, installation, upgrading or repair of the device
may only be carried out by specially trained and in-
structed personnel.
Information regarding the disassembly and assembly of the device is avail-
able in the separate assembly instructions. These assembly instructions are
intended solely for use by the manufacturer's service personnel or by qualified
specialists authorized by the manufacturer.
2.6.2
Ventilation
WARNING!
Risk of overheating due to air flow problems!
Operating faults possible.
–
Do not change, obstruct or block the air inflow and
outflow to the device or make any unauthorized
changes to the device. The manufacturer shall bear no
liability for damage or injuries resulting from such ac-
tions.
–
Observe the ventilation system data.
–
Do not set up and operate the device on the transport
pallet!
Data regarding the air supply and ventilation planning is available in the plan-
ning manual (www.ergoline.de, available in English and German). You can
also obtain this data from your dealership or the customer service department
(see page 2).