- 16 -
ergoline
GmbH
ergoline
GmbH
o
Peration
The ergometers of the ergoselect series are available with
two versions of the control terminal whose functionalities
differ.
The following sections describe the control and configura‑
tion of the ergometer.
C
ontrol
terminal
P
t
urninG
the
s
ystem
o
n
You turn the ergometer on by pressing the power switch ‑
the green indicator in the switch lights up.
The ergometer runs a self‑test. Subsequently, the main
menu displays.
Control terminal P
Control terminal K
Note
•
Instruct the patient not to pedal while the ergometer is
being turned on and during the self-test.
•
Apply the blood pressure cuff to the patient AFTER the
ergometer has been turned on and the self-test completed.
•
The device can be configured to default to one of the
operating modes.
If this option is selected, the initial screen of the selected
operating mode (e.g. Ergometry) will be displayed instead
of the main menu.
With the
key, you can display the main menu.
•
If an error message (e.g. E:01) appears immediately after
the self-test, please refer to the "Troubleshooting" section
for advice.
s
elf
-
test
sCreen
Selftest running
m
ain
menu
PC Mode
Ergometry
Manual
Settings
Select
The ergometer software is controlled with 5 keys:
With this key you display the main menu or
return to the previous menu level.
With this key you initiate a blood pressure mea‑
surement. A measurement in progress can be aborted with
the same key.
The functions of these three softkeys
change with the displayed menu ‑ the key label describing
the function is shown on the display.
k
eypaD
p