ergolet Luna IPx4 User Instruction Download Page 10

10 

2.3 Pictograms and functions

 

(1) ON

 – If LED is off Luna is in “sleep” mode. 

The LED light will turn on when the button up 

(4)

 or down 

(3) 

on the hand control is pressed.

 

Side foil

(2) BATT.

 – Indicates the level of the battery 

(see section 2.6).

 

(3)

+

(4)

 

Up/Down –

 With a constant hold the 

Luna will move in the desired direction. 
Release the button to stop the movement. 

Note

: Works only with hand control 

(5)

 

plugged in and LED 

(1)

 must be on.

 

(5)

 

Plug 

– Insert the hand control plug.

 

 
 

Hand control

:

  

(6) Data Plate – 

Information about the Luna 

model, serial number and symbols 
information.

 

(7)

 

Emergency Lowering

 – ONLY used in 

situations where it is necessary to lower a 
client safely in a lift failure. This button must 
be held constantly to lower the unit. 
(See section 2.5) 

Note

: Works only with hand control 

(5) 

plugged in and LED 

(1)

 must be on.

 

(8)

 

STOP 

– Press the emergency stop button to 

stop the Luna immediately. This function can 
be reset by pressing the reset button on the 
hand control for 5 seconds.

 

9) Reset 

– Used to reset the Luna after the use 

of the emergency stop - or lock function. Reset 
by holding the reset button for 5 seconds, the 
Luna will them be reactivated. 
Reset can also be used to lock all functions to 
avoid unintended use by e.g. the client. When 
the reset button is held, the LED light will 
show red and all functions will be locked. 
Deactivate as after an emergency stop 

(10) Option – 

This option is not available on a 

standard hand control. The extra output can 
be for e.g. an electrical spreader bar.

 

 

 

(1)  (2) 

(3) 

(4) 

(5) 

(6) 

(7)  (8) 

(9) 

(7) 

(8) 

(10) 

(4) 

(3) 

(2) 

(1) 

Summary of Contents for Luna IPx4

Page 1: ...must be read before the product is used This user instruction is intended for technicians installers therapists and other personnel in connection with the use servicing and maintenance of the LunaIPx...

Page 2: ...2...

Page 3: ...4 Operation 11 2 5 Operation of safety functions 11 2 6 Charging 13 3 Using Luna 3 1 Using lifting slings 14 3 2 Room to room transfer 15 3 3 Using Luna in a shower environment 17 4 Maintenance 4 1 C...

Page 4: ...rmal temperatures 2 to 40 C and relative humidity 40 90 If Luna is used near shower systems or bath it should be ensured that the Luna or hand control is not submerged see section 3 3 Luna should neve...

Page 5: ...ied technician When the up down function are checked ensure at the same time that the strap not folds or twists Check also that the strap moves freely into the Luna Lifting slings are checked for any...

Page 6: ...t in case of repair or annual check Contents of the box Luna cassette User manual Hand control accessories Spreader bar accessories Charger accessories 1 6 Transport The Luna ceiling hoist is most saf...

Page 7: ...d connect the bayonet bracket into the Luna resistance will be felt since the bracket is spring loaded 2 Luna is now turned 90 clockwise When Luna is in place a click will be heard This indicates corr...

Page 8: ...olley with extension arm can be purchased from Ergolet or your local dealer See order number in section 8 Accessories 2 2 Mounting and detaching the spreader bar Depending on the Luna cassette mountin...

Page 9: ...t bracket on the Luna cassette Resistance will be felt due to a spring load 2 Turn the spreader bar 90 clockwise See fig 2 2B When the spreader bar is in place a click will be heard indicating correct...

Page 10: ...ft failure This button must be held constantly to lower the unit See section 2 5 Note Works only with hand control 5 plugged in and LED 1 must be on 8 STOP Press the emergency stop button to stop the...

Page 11: ...unctions will stop immediately and the LED light will show red on the hand control The Emergency stop can be deactivated by pressing the Reset button on the handset for 5 seconds Emergency lowering bu...

Page 12: ...1 Remove the hand control Fig 2 Remove the plug on the side of the Luna Fig 3 Take the key which is placed at the top of the charger Put the key into the Luna and turn the key anti clockwise to lower...

Page 13: ...rmation concerning battery condition during use Diode color Description Green The LED on the hand control will flash during charging The LED lights constantly on the hand control when the batteries ar...

Page 14: ...d also the correct use of slings Hanging slings on the Ergolet spreader bar Ergolet accepts no responsibility if slings are placed incorrectly it is always the onus of the carer helper to ensure that...

Page 15: ...after which the lifting strap from Luna b is extended and fitted on the extra bracket ring on the spreader bar 3 Now with both Luna cassettes connected to the spreader bar Check that the hook is fitte...

Page 16: ...ps b carries user weight 5 The strap on Luna a can now be removed and fitted on the cart next to the transfer strap The lifting strap is best removed by bringing the bracket up press the spring and re...

Page 17: ...immersed in water 4 MAINTENANCE 4 1 Cleaning Luna may be cleaned using a slightly damp cloth with a small quantity of standard household cleaning agent Strong chemicals or abrasive agents must not be...

Page 18: ...more information contact Ergolet 5 2 Troubleshooting Fault Possible cause Solution Luna does not react when the hand control UP DOWN are used No beep sound from the Luna unit Luna is not on or emerge...

Page 19: ...Try another charger or get a new one from the dealer or Ergolet Connection to the hand control is defective Try another charger or get a new one from the dealer or Ergolet 6 TECHNICAL DATA AND DIMENSI...

Page 20: ...eight of Luna cassette 170 mm D Min height from hook to track 430 mm E Lifting range from min height 1950 mm Measurements strap mounted A Width of Luna cassette 320 mm B Spreader bar see section 8 acc...

Page 21: ...gh reuse or recycling Luna is equipped with 1 battery which may be returned either to the local recycling station or to Ergolet at the end of their lifetime Fractions that contain aluminium lead acid...

Page 22: ...bols and abbreviations Direct current Double insulated Type B patient is not separated from the ground Type B patient is not separated from the ground Refer to user instruction Safe working load user...

Page 23: ...rap 58 cm Spreader bar 200 kg Bayonet 58 cm Spreader bar 200 kg Bayonet 48 cm Spreader bar 200 kg Bayonet 36 cm Spreader bar 275 kg Bayonet 58 cm Spreader bar with double hook for room to room 015 019...

Page 24: ...24 Manufacturer Ergolet Taarnborgvej 12 C 4220 Korsoer Denmark Phone 45 70 27 37 20 Fax 45 70 27 37 19 Email ergolet ergolet dk www ergolet com 93 42 E F CLASS I EN ISO 10535 EN 60601 1 2 910 00766UK...

Reviews: