background image

 

 

X K 3 1 9 0 - A 1

+P 

 

 

 

The printer must be compatible to the indicator, please choose the recommended 

printer. 

 

The printer GND can not be connected to the power GND, otherwise the indicator 

or printer will be destroyed. 

  3. Set print parameters 

 

step operating 

display 

Explanation 

press [PRN Set ] 

press [ 9 ] [ 7 ] 

press [ Enter ] 

[ P   00 ] 

[ P   97 ] 

Enter printing setup 

Input password 97 

2  

 

 

press [ 0 ] 

press [ Enter ]   

[ Auto  1] 

 

 

[ Auto  0 ] 

Auto/ manual print 

0: manual 

1: Auto 

Example: 0 

3  

 

 

 

press [ 2 ] 

press [ Enter ] 

[ type   0 ] 

 

 

 

[ type   2 ] 

Selecting type of the printer. 

0: print inhibited 

1: TPup16TP (Eng. mode) 

2: TIMES TM-800 

3 National KX-P1121 

4: EPSON LQ-1600K 

Example: 2 

4  

 

 

press [ 0 ] 

press [ Enter ] 

[ H-u   1 ] 

 

 

[ H-u   0 ] 

Selecting print format 

0: Record format 

1: document format 

Example:0 

5  

 

press [ 3 ] 

press [ Enter ] 

[ Arr   1 ] 

 

[ Arr   3 ] 

selecting document numbers   

(1-3) 

Example: 3 

6  

 

press [ 1 ][ 0 ][ 0 ] 

press [ Enter ] 

[ L 000.10 ] 

 

[ L001.00 ] 

 

Selecting Auto-print lower 

Limit. 

  L: must> 10e 

Example: 1:00 

7  

Press[0] 

Press[Enter] 

[ ode    *] 

[ ode          ] 

0: doesn’t choose blank 

filling mode 

1: choose blank filling print 

mode 

Press[0] 

[ dct    *]  0: doesn’t use discount 

Summary of Contents for Weighing Indicator XK3190-A1+P

Page 1: ...XK3190 A1 P Weighing Indicator USER S GUIDE Attention Please read this User s Guide carefully when you use the indicator ...

Page 2: ...mmunication port to Indicator 4 7 Setting of Time and Date 4 8 Storing Checking and Deleting of Data 4 9 insufficient voltage indication 5 Maintenance and Announcements 6 Errors and Information Important Notes 1 Printer interface invalid due to the built in micro printer optional supply upon special requirement 2 The function key is used as the paper roll key it works upon pressing and stops worki...

Page 3: ...rface Using serial output method current loop signal transmission distance 50m RS323 signal transmission distance 30m 8 Communication port RS 232C Baud rate 600 1200 2400 4800 9600 optional 9 Printing Port Standard parallel output port can connect with Tpup16 micro printer TM800 LX 300 KXP 1121and LQ 1600k wide line printer 10 Power Supply AC 187 242V 49 50Hz The built in repair proof storage cell...

Page 4: ...RE WEIGH 7 SET PRINT 8 CALB 9 TIME PRINT ZERO UNIT KG MONEY PRICE AUTO DATE TIME TARE STABLE Graph 2 1 Front View of Indicator ACCU accumulating FUN function CALB Calibrating 25 pins 15 pins 9 pins Print output port RS 232C Scoreboard output port Loadcell input port Nameplate Power source Fuse Graph 2 2 Back View of the Indicator ...

Page 5: ... inputs digital key 8 8 TIME 9 In calibration mode pressing TIME 9 can input digital key 9 In weighing mode it is time function key more details shown in episode setting of time and date 9 STORE 0 In calibration mode pressing STORE 0 can input digital key 0 In weighing mode it is storing function key more details shown in episode storing data 10 ACCU 1 UNIT PRICE 3 In calibration mode pressing ACC...

Page 6: ...must be reliably connect ed to Loadcell and shielded cable of loadcell must be reliably connect ed to underground If indicator is powered on the user should not insert or withdraw the plug in order to protect the indicator and loadcell 4 Sensor and indicator are static sensitive devices you must adopt anti static measures In order to protect the operator indicator and relevant devices you should i...

Page 7: ...ecting 1 2 5 10 20 50 100 200 Example 10 3 press 0 press Enter dc dc 0 Enter decimal point 0 4 Example without Decimal point 0 4 press 2 3 press Enter pon XY pon 23 Enter Zero range X zero set range 1 5 it stands for 2 4 10 20 100 of F S Y power on Zero set range it Stands for 1 5 2 4 10 20 100 of F S example 23 5 press 3 0 0 9 0 F F 30090 Enter overload warning value when inputting F value and re...

Page 8: ...H L td indicating three calibrating coefficients Recording them for spare use 12 press 1 press Enter Adr Adr 01 communication address 01 26 Example 1 13 press 1 press Enter bt bt 1 Serial communication baud rate 0 4 It respectively represents baud rate 600 1200 2400 4800 9600 Example 1 1200 14 press 0 press Enter tF tF 0 Serial communication Mode 0 Continuous transmitting mode no receiving 1 Comma...

Page 9: ...setting 1 In weighing mode press Zero key Indicator will perform zero setting 2 If displayed value deviates from zero point but still within zero range pressing Zero key is available Otherwise Zero key is invalid In this status please recalibrate or reset zero parameters 3 Only when stability sign lamp is on you are allowed to set zero operation 4 3 Tare function 1 Normal Tare When Indicator at we...

Page 10: ...6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 ST D0 D1 D2 D3 D4 D5 D6 D7 BUSY GND 2 Printing notes You must set the printer before you use it You must connect the indicator and the printer correctly The wrong connection will destroy indicator output interface or printer input interface and may even destroy the indicator or printer Turning on the printer as following connecting the line c...

Page 11: ...int 0 manual 1 Auto Example 0 3 press 2 press Enter type 0 type 2 Selecting type of the printer 0 print inhibited 1 TPup16TP Eng mode 2 TIMES TM 800 3 National KX P1121 4 EPSON LQ 1600K Example 2 4 press 0 press Enter H u 1 H u 0 Selecting print format 0 Record format 1 document format Example 0 5 press 3 press Enter Arr 1 Arr 3 selecting document numbers 1 3 Example 3 6 press 1 0 0 press Enter L ...

Page 12: ...ross kg Tare kg Net kg Accumulating kg 0001 08 56 16 299 98 1 298 98 298 98 0002 09 00 09 299 98 2 297 98 596 96 0003 09 00 28 299 98 3 296 98 893 94 Document Format Weight Bill NO 007 Date 96 09 25 Time 09 03 21 Gross 299 98 kg Tare 9 98 kg Net 290 00 kg Total 2059 90 kg Blank filling format Compelet printing only in five sconds WEIGHT BILL Fist bill for operator SERIAL No 123 DATE 1999 05 28 TIM...

Page 13: ...t 4 6 Connecting Serial Communication port to Indicator 1 Communication port adopts 15 pins socket which was shared with scoreboard It was shown on page 9 of User s Guide for A1 the pin 6 RXD is used to receive the signal from the computer the pin 7 TXD is used to output the signal to computer 2 Communication port adopts RS 232C and all data are ASCII codes Every group of data consists of ten bits...

Page 14: ...Don t change this data and enter next step 5 Press Enter F Don t change this data and enter next step 6 Press Enter Press Enter H L td Don t change this data and enter next step 7 Press 1 Press Enter A d r A d r01 Communication address 01 26 Example 1 8 Press 1 Press Enter bt bt 1 Baud rate of serial communication 0 4 which represent baud rate of 600 1200 2400 4800 9600 9 Press 0 Press Enter tF tF...

Page 15: ...total accumulated times total accumulated weight value and 200 groups of weighing records 5 When print control Auto is set to 1 system not only auto prints but also auto stories and auto accumulates weighing data 2 Checking Press Check key to check the weight record data in internal memory of the Indicator Including Total accumulating times Total accumulating weight value and every group of weighi...

Page 16: ...nce and Announcements 5 1 To guarantee the clarity and using life the indicator shouldn t be placed directly under sunshine and should be placed in the plain space 5 2 The indicator should avoid dust pollution vibration and moisture 5 3 Loadcell should connect with indicator reliably and the system should be connected into ground properly The indicator must be detected from high electrical fields ...

Page 17: ...his indicator to our company for repair Non weighing manufacturer should not repair it or by yourself otherwise further destruction may occur 5 8 From invoice date the indicator has a total one year free repair period If any non artificially obstacle about the indicator occurs while under correct using conditions within the period the user is allowed to send the product with its guarantee card of ...

Page 18: ...till indicates Err01 the problem can then be solved by short connecting pin 4 and pin 6 3 Err 02 It means the load cell s wires are wrongly linked or the signal value exceeds the A D converting range 1 If this scale is under usage then can be sure the load cell s connective wires had troubles or load cell is damaged 2 If this scale hasn t been calibrated please check as following a Carefully check...

Page 19: ...ouldn t be put in If calibration is needed the calibration jumper should be put in at set up You should pull out the calibration jump when turning off the indicator If the indicator finds that the calibration jump was inserted when turning on it will display Err 24 for 6 seconds to prompt the operator 2 Err 25 Illegal software or E2PROM was damaged 12 groups of date constitutes of a frame Date No ...

Page 20: ... the computer consists of 6 data explanation 1 02 X ON start 2 A Z address 3 A Shaking hands B asking GW C asking tare D asking NW E asking accumulated times and accumulated weight 4 XRL checking 5 XRL checking 6 03 X OFF end XRL 2 3 ...

Page 21: ... 6 groups of data Command B GW GW data 6 bits GW data Decimal position Each frame consists of 14 groups of data Command C Tare Tare data Tare data Decimal position Each frame consists of 14 groups of data Command D NW NW data NW data Decimal position Each frame consists of 14 groups of data ...

Page 22: ...X K 3 1 9 0 A 1 P Command E Accumulated times and weigh Accumulated times 4 bits comma separation mark Accumulated weight 10 bits Decimal position Each frame consists of 22 groups of data ...