Maintenance
13
Maintenance
Cleaning the Associate
Parts for
ILS20EP/ILS20EPUV
If the color measurement value indicates
abnormal value, clean the product. To maintain a
good condition of the product, we recommend
you to care for once a month.
Cleaning procedures
Clean the following three parts:
❏
the backing surface
❏
the surface of the white calibration tile
❏
the lens of ILS20EP/ILS20EPUV
The procedures are as follows.
A
Remove ILS20EP/ILS20EPUV and the
mounter from the printer.
See “Removing the Mounter” on page 15
B
Remove the white calibration tile holder
from the printer.
C
Remove the white calibration tile from the
white calibration tile holder.
D
Remove the backing from the printer.
E
Use a dry, soft cloth to carefully remove any
dust or dirt from the surface of the white
calibration tile or the surface of the backing.
If the dirt does not come off, clean it using a
soft, clean cloth dampened with water.
Then, wipe it with a dry, soft cloth.
c
Caution:
Because the backing is a fragile part, do
not drop the backing while removing it.