4.3.3.2 Actual values
Actual values of output voltage (with and without remote sensing) and output current are calibrated almost the
same way as the set values, but here you don’t need to enter anything, just confirm the displayed values. Please
proceed the above steps and instead of “
Output value
” select “
Actual value
” in the submenus. After the device
shows measured values on display, wait at least 2s for measured value to settle and then simply confirm with
, until you are through all steps.
4.3.3.3 Saving calibration data
After calibration you can furthermore enter the current date. To do so, navigate to menu item “
Cal. date
” and enter
the date in format YYYY / MM / DD and submit with
.
Last but not least save the calibration data permanently confirming menu item “
Save & exit
” with
.
Leaving the calibration selection menu without saving via “Save & exit” will discard
calibration data and the procedure would have to be repeated!
5.
Contact and support
5.1
General
Repairs, if not otherwise arranged between supplier and customer, will be carried out by the manufacturer. For this
the device must generally be returned to the manufacturer. No RMA number is needed. It is sufficient to package
the equipment adequately and send it, together with a detailed description of the fault and, if still under guarantee,
a copy of the invoice, to the following address.
5.2
Contact options
Questions or problems with operation of the device, use of optional components, with the documentation or soft-
ware, can be addressed to technical support either by telephone or e-Mail.