INFORMATION ABOUT TUNING AND CALIBRATION
The resulting calibration coefficients (
a, b, c, d, e, f
) are pointed in the technical
report provided with your MDS-
4 module.
In order to treat the signals we recommend using a DAQ device
and a PC for signal
processing using the formulas above with calibration coefficients.
We recommend performing calibration c
hecks annually. Se
nsor module
recalibration procedure involves the adjustment of “
a
” coefficient only, all other
coefficients remain unchanged.
To define the recalibrated “
a
” coefficient
value one
needs to measure
U
SD
and
U
T
at a specific
temperature with
a
zero methane level and
use the following formula:
𝒂
𝒓𝒆𝒄𝒂𝒍𝒊𝒃𝒓𝒂𝒕𝒆𝒅
=
𝑈
��
−
(
𝒃
×
𝑈
�
+
𝒄
×
𝑈
�
�
)