Application & Description 

 

MDS-4 is a 

methane sensor module for

 CH

4

 

detection

. It 

includes a compact optical 

cell and electronics for LED

 

power supply and PD signal amplification all

-in-one. 

 

Features 

 

Diffusion type detection

 

 

Measurement in 0

-5% 

(volume)

 concentration range

 

 

Resolution down to 

250 

ppm in 0

-5% 

(volume) 

concentration range 

 

 

Very low power consumption 

– 

3.5 mW

 

 

 

Quick 

response time

 – <2 s

 

 

Operating t

emperature 

range – 0..+40°C 

 

 

Size: 55x26x14 mm (including optical cell and circuitry)

 

 

Gases: pre

calibrated for

 

methane, but will respond to most hydrocarbons

 

 

Possibility 

of

 

integration  with  wireless  data  transfer  protocols  like  Zigbee,  WiFi, 

GPRS 

 

Possibility 

of

 power battery supply  

 

Appearance & Layout

 

 

1.

 

Optical cell with an LED 

Lms34LED

-

CG and a photodiode Lms36PD

-05-CG 

2.

 

Power input 

3.

 

Temperature and m

easuring signal output 

 

Temperature compensation

 

In t

he  module

  there  is  realis

ed  a  circuit  for  measurement  of  LED’s

  operation 

voltage  for  temperature  determination  of  the  optical  cell

.  It 

enables  temperature 

compensation

 

of measuring signal 

in 0...+40°C range. 

 

Operation conditions

 

Indoor operation only.

 Ingress Protection Rating IP00. 

 

 

 

 

Summary of Contents for MDS-4

Page 1: ...Electro Optical Components Inc 5464 Skylane Boulevard Suite D Santa Rosa CA 95403 Toll Free 855 EOC 6300 www eoc inc com info eoc inc com MDS 4 METHANE SENSOR MODULE INSTRUCTION MANUAL rev 090915...

Page 2: ...Application Description 3 Features 3 Temperature compensation 3 Operation conditions 3 Information about tuning and calibration 4 5 Operating instructions 6 Precautions 6 Appendix MDS 4 testing resul...

Page 3: ...ical cell and circuitry Gases precalibrated for methane but will respond to most hydrocarbons Possibility of integration with wireless data transfer protocols like Zigbee WiFi GPRS Possibility of powe...

Page 4: ...ature change from 0 C to 40 C range with 0 2 C min During this procedure signal values from UT temperature signal and USD measuring signal are measured with 0 04 C resolution Basing on this data array...

Page 5: ...and a PC for signal processing using the formulas above with calibration coefficients We recommend performing calibration checks annually Sensor module recalibration procedure involves the adjustment...

Page 6: ...ate to your MDS 4 module Precautions Turn on the power supply only after all connections are made and tested Do not disassemble the optical cell otherwise the optical system will be damaged Do not use...

Page 7: ...eviation mV Noise mV Resolution ppm Air Dry Bottled 0 2 110 2 17 7 58 CH4 N2 1000 2 119 252 CH4 N2 2500 2 136 237 CH4 N2 5000 2 158 253 CH4 N2 10000 2 190 303 CH4 N2 50000 2 304 632 2 11 2 16 2 21 2 2...

Page 8: ...8 APPENDIX Relative signal change dependence on methane concentration USD signal output 8 7 6 5 4 3 2 1 0 0 1 2 3 4 5 Relative signal change Methane concentration...