MT-10 + MT-5 STERILISATION
Sterilisation has to be done when the content of bacteria is too high, which you recognise
on a bad taste of the water. Pretreat the unit with MT-10 as described in ”MT-10
CHEMICAL CLEANING”. Flush for at least 15 minutes. After that you are going to treat
the unit as described in ”Preparing MT-5 Sterilisation”.
Prepare MT-5 Sterilisation
Mix one plastic bottles of MT-5 into the 40 litres cleaning tank containing product water.
Mix well.
Re-circulation time MT-5 Sterilisation
1 hour.
If the unit is intended to be in service you must proceed as described in ”INITIAL
START-UP PROCEDURE”.
If the unit is intended to be stored you must leave it with the MT-5 solution.
Do not flush the unit then.
MT-5 PRESERVATION
This solution shall be used when the unit will be out of service for more then 1 week and
you are not intended to flush it every week.
Prepare MT-5 Preservation
Add one plastic bottle of MT-5 to the 40 litres cleaning tank containing product water.
Mix well.
NOTE
1:
The MT-5 solution must be in the unit during the preservation time (max. 1 year) so you
must not flush it after preservation. When the unit is intended to be in service again you must
proceed as described in ”INITIAL START-UP PROCEDURE”.
NOTE 2
:
The unit is now preserved with MT-5 solution for 1 year. If the unit will be stopped more
than 1 year it must be re-preserved.