General set up
41
P 3.1 Auto Stable, Auto Hold and Real Time mode setup
Auto Stable (AS) mode
- the meter shows live readings
annunciator blinks until the reading
is stable.
Auto Hold (AH) mode
- the meter locks the stable reading;
annunciator blinks until reading is
stable and then
lights up.
Real Time (RT) mode
- the meter shows live readings; Both
and
annunciators are
inactive.
9. Press the
key,
P1 PH/DO
screen appears.
10. Press
key,
P2 DATA
screen appears.
11. Press
key,
P3 GEN
screen appears.
12. Press the
ENT
key,
P3.1 STBL
screen appears.
13. Press the
ENT
key, Default the stability mode is
AS
(auto stable).
14. Use the
keys to change the stability mode as
AH
(auto hold) or
RT
(real time).
15. Press the
ENT
key,
P3.1 STBL
screen appears. This indicates completion of the
stability mode selection.