ORP calibration
14
Calibration
11. After placing the electrode in the solution, press the
key to switch to mV mode.
12. Press the
key.
13. Meter starts reading mV values and the
blinks until value stabilizes.
14. Wait for the
to stabilize (stable calibration reading).
15. Use the
keys to adjust the mV value to your desired value.
16. Press the
ENT
key to confirm and save calibration data.
17. Meter displays
DONE
that indicates end of the ORP/mV calibration procedure.