background image

+1.800.343.5319 or +1.843.881.6560    envexp.com    

 

Environmental Express 

• 7

HotBlock

® 

TKN:

 Operation & Instruction Manual  

 

 

 

Care & Maintenance

Maintenance

Any service inquiries should be directed to Environmental Express Technical Service 
Department. 
After each use, clean exterior surfaces with a damp sponge to remove acid residue. 
For acid spills, sponge with a diluted solution of sodium bicarbonate followed 
by distilled water. Acid that is spilled directly into the digestion wells should be 
neutralized and removed. 
Before using any cleaning or decontamination methods except those 
recommended, check with Environmental Express to confirm the proposed method 
will not damage your HotBlock® TKN unit. 
Avoid excessive spills, as liquid allowed to overflow into the HotBlock® TKN casing 
can severely damage electronic components.
The controller is independent from the block and should be placed outside the 
fume hood to reduce the likelihood of acid damage.

Troubleshooting

The HotBlock® TKN contains no user serviceable parts. If you experience troubles 
with heating, temperature accuracy, controller display, or any other issues, you 
will need to send the affected portion of your system to Environmental Express for 
repairs. Please contact one of our technical support staff to coordinate the repair 
process.
To identify the defective area you will need to take the resistance of the heating 
mat and RTD. This is done through the cord that connects the block to the 
controller. Pins 1 and 2 are connected to the heating mat. For a 54-well block the 
resistance should be 16 Ω. For a 28-well block the resistance should be 20 Ω.  
Pins 6 and 7 are connected to the RTD. The resistance should be 110 Ω. If these 
readings are correct the problems lies within the controller. 

Summary of Contents for HotBlock TKN System

Page 1: ...HotBlock TKN System Operation Instruction Manual 2021 Environmental Express Inc All rights reserved July 2021 1065EE1_M Rev 4 1 800 343 5319 or 1 843 881 6560 envexp com...

Page 2: ...ing Your HotBlock TKN 6 Safe Sample Temperature Protection 6 Care Maintenance Maintenance 7 Troubleshooting 7 Circuitry Diagram 8 Replacement Parts Supplies 9 Product Information Item Date of Purchase...

Page 3: ...isclaims all other warranties Environmental Express is not liable for any consequential special indirect or compensatory damages arising from use of or in conjunction with this product the maximum lia...

Page 4: ...grounding pin See electrical requirements for power specifications page 4 Application of the wrong supply voltage can create a fire hazard and a potentially serious shock hazard and could seriously d...

Page 5: ...ge alternating current Hz frequency A amperes Unpacking Your HotBlock TKN System 1 Remove the HotBlock TKN from the shipping container by lifting from the bottom of the block 2 If you purchased a comp...

Page 6: ...e range 5 to 30 C Ambient relative humidity 0 to 90 RH Altitude sea level to 2500 meters HotBlock TKN is rated as Pollution Degree 2 and Installation Category 2 Electrical Requirements Required Voltag...

Page 7: ...ion area Advanced Touch Screen Controller 1 Turn on the controller wait for the opening screen to appear 2 Touch the screen to proceed to the Main Menu Screen 3 Select the Global Settings menu 4 Touch...

Page 8: ...does not have active cooling and cannot reliably use a time function to go down in temperature set point b Soak Use to hold the block at a temperature for a specific length of time c End Signifies the...

Page 9: ...g can severely damage electronic components The controller is independent from the block and should be placed outside the fume hood to reduce the likelihood of acid damage Troubleshooting The HotBlock...

Page 10: ...1 800 343 5319 or 1 843 881 6560 envexp com Environmental Express 8 HotBlock TKN Operation Instruction Manual Circuitry Diagram...

Page 11: ...rts Supplies Description Item Advanced Touch Screen Controller TKN180 Simple Manual Controller TKN100 Replacement RTD Sensor TKN942 Replacement Heater Mat 54 Well Block TKN966 Replacement Heater Mat 2...

Reviews: