background image

 

- 9 - 

4.3.3

 

The 3

rd

 point calibration 

Repeat  the  steps  as  per  Section  4.3.2,  use  pH10.01  buffer  solution,  and  complete  the  3

rd

  point 

calibration. The bottom left of LCD indicates                      .

 

 

4.4 

Sample test

 

Rinse pH electrode in distilled or deionized water, allow it to dry, and submerge it in the sample 
solution. Stir the solution briefly and allow it to stay in the sample until the value becomes stable 
and          icon appears on LCD, get the reading which is the pH value of the sample solution. 

 

 

4.5 

Information regarding pH calibration and measurement 

 

4.5.1

 

During the process of calibration and measurement, submerge both the pH electrode and 

the temperature probe into the solution to ensure automatic temperature compensation. 

 

4.5.2

 

In calibration mode, press          to exit.   

 

4.5.3

 

The meter can conduct 1~3 points of calibration according to measurement requirements. 

Refer to Table - 3 to check when to adopt. 

 

4.5.4

 

During the process of calibration and measurement, the meter has self-diagnosis functions, 

indicating information Er1~Er4, please refer to Table - 4 for relevant solutions. 

 

4.5.5

 

   

The calibration frequency depends on the sample, electrode performance and accuracy

 

requirement. For high accuracy measurements (≤±0.03pH), the meter should be calibrated before 
testing every day. For ordinary accuracy measurements (≤±0.1pH), once calibrated, the meter can 
be used for about a week or even longer. 

 

 

In the following cases, the meter must be re-calibrated: 

a)

 

The electrode hasn’t been used for a long time or the electrode is brand new.

 

b)

 

After measuring strong acid (pH<2) or strong base (pH>12) solutions.

 

c)

 

After measuring fluoride-containing solution and strong organic solution.

 

d)

 

There is a big difference between the temperature of the test sample and the temperature 

of the buffer solution that is used in the last calibration.

 

4.5.6

 

pH temperature principle 

Please note that the closer the temperature of the sample is to the calibration solution, the more 
accurate will be the readings. 

 

Summary of Contents for 35613-65

Page 1: ...ION 100 Portable Ion Meter Instruction Manual 35613 65 68 Environmental Express www environmentalexpress com...

Page 2: ...nformation regarding pH calibration and measurement 9 4 6 Customized calibration take 1 60pH 6 50pH as example 10 4 7 Notes regarding customized calibration 10 4 8 Self diagnosis information 10 4 9 Ma...

Page 3: ...der to use and maintain the instrument properly please read the manual thoroughly before use To improve instrument performance constantly we reserve the right to change the manual and accessories with...

Page 4: ...86 9 18 and 12 45 pH NIST mV Range 1000mV Resolution 200 200mV 0 1mV Other range 1mV Accuracy 0 2 F S Ion Range pX 0 00 10 00 Ion concentration 0 1999 Switchable unit pX mol L ppm mg L Resolution 3 4...

Page 5: ...g ATC automatic temperature compensation MTC manual temperature compensation Ion symbol Calibration guide icon Units of measurement Automatic reading hold icon Temperature units Stability icon of read...

Page 6: ...rt press In measurement mode press to turn on off the backlight In calibration mode press to calibrate In the parameter setup mode press to enter and open the sub menu confirm parameter When measureme...

Page 7: ...to change the temperature value long press for quick changing of the temperature value 3 5 2 Automatic Power off The meter is with automatic power off function Select from parameter P6 4 to select aut...

Page 8: ...rectly goes to automatic temperature compensation mode When the meter is connected to the pH electrode only it will enter manual temperature compensation mode Press or to input the temperature into th...

Page 9: ...H7 00 buffer to conduct the 1st point of calibration Dip the pH electrode and the temperature probe into pH7 00 buffer solution stir gently and let it stand still and wait for the reading to be stable...

Page 10: ...adopt 4 5 4 During the process of calibration and measurement the meter has self diagnosis functions indicating information Er1 Er4 please refer to Table 4 for relevant solutions 4 5 5 The calibratio...

Page 11: ...into 1 60pH calibration solution stir gently and let it stand still and wait for the reading to become stable When the smiley icon displays on LCD press to adjust the value to 1 60 then press to confi...

Page 12: ...ce The soaking solution contained in the supplied protective bottle is used to maintain hydration of the glass bulb and liquid junction Loose the cap take out the electrode and rinse in distilled or d...

Page 13: ...ss bulb Electrodes that have been used for a long period of time will become aged Submerge the electrode in 0 1 mol L hydrochloric acid for 24 hours then rinse the electrode in distilled water then su...

Page 14: ...quality or measuring value use ORP standard solution to test mV and see whether ORP electrode or meter works properly 5 3 2 Clean and activate ORP electrode After the electrode has been used over a lo...

Page 15: ...ing should be 1 00 x 10 3 mol L When we use the unit of pX and ppm the reading we get is the value see Diagram 7 and 9 6 1 3 Choose the ion type The meter is built in with 16 ions frequently being use...

Page 16: ...m right corner of the LCD When the reading is stable and the stable ion displays on LCD press or to adjust the value to be 1 00 E 5 1 00 10 5 press to finish the 1st point calibration the bottom left...

Page 17: ...o be rinsed and the solution contamination can be avoided 6 3 3 Calibration error reminder Ion electrode calibration may fail often The reason could be calibration solution error the quality problem o...

Page 18: ...4 Factory default setting For factory default setting please refer to parameter setting P5 3 All calibration data is deleted and the meter restores to the theory value Some functions restore to the or...

Page 19: ...n menu and P6 0 for basic parameter menu Diagram 13 Main menu of parameter setting Sub menu of pH parameter setting Sub menu of ion parameter setting Sub menu of basic parameter setting P1 1 pH buffer...

Page 20: ...ss to confirm USA USAseries nIS NIST series CUS Customized calibration 2 Press to enter P1 2 or press toreturn P1 2 Resolution 0 01 0 1 1 Press and 0 01 flashes press to select between 0 01 0 1 Press...

Page 21: ...er the ion selection interface see Diagram b 2 Press or to move the cursor press to confirm 3 Press to enter P5 2 or press to return P5 2 Select between ppm and mg L 1 Press ppm flashes press to selec...

Page 22: ...press to confirm Off Unlock up On lock up Stable reading 10 seconds and the reading will be automatically locked up 2 Press to enter P6 3 or press to return P6 3 Select backlight timing 1 2 3 Off 1 P...

Page 23: ...ttributable to responsibility of Environmental Express for a period of THREE YEARS SIX MONTHS for the probe from the delivery This limited warranty does not cover any damages due to Transportation sto...

Page 24: ...e to factory default setting No Yes No P6 0 Basic Parameter P6 1 Temperature unit P6 2 Automatic lock up timing Off On P6 3 Backlight timing 1 2 3 Off P6 4 Automatic power off timing 10 20 30 Off Appe...

Page 25: ...ort 800 745 8218 Email tech environmentalexpress com Website www environmentalexpress com Icon Self diagnosis Information pH Ion Press when measuring value is not stable during calibration Wrong pH bu...

Reviews: