Error Code 

Description 

Recommended Action 

E5 

 Unknown 

Test 

QuickScan did not 
recognize the test kit, 
and a software upgrade 
is probably required.  

Check that the test strips have been 
placed correctly in the tray, and read 
again. If error message reoccurs, 
contact EnviroLogix Tech Service.  

Various codes for 
control lines, test 
lines and 
background 

Imaging software 
cannot read the strips 
properly. Flow of liquid 
sample up the test strip 
may have been 
impeded or failed. 

Examine the test strips for proper 
flow. Check that the strips have been 
placed correctly on the tray. Consult 
test kit Product Insert for proper kit 
operation, as new test may need to be 
performed. 

E20 

 Printer not 

activated 

A printer has not been 
selected in the QS 
configuration window, 
or intended printer not 
been installed.  

Make sure the intended printer has 
been installed on the QuickScan, then 
select that printer in the QuickScan 
configuration window. 

E30 

 Outside 

image border 

Test strips have been 
placed in tray 
incorrectly. 

Reposition strips and read again. 

 

DIMENSIONS AND WEIGHT 

Length: 37 cm 
Width: 33.5 cm 
Height: 20 cm 

 

Weight:  11.15 lbs / 5.06 kg 

 

SAFETY AND PRECAUTIONS 

1.

 

QuickScan is a precision system that combines a digital imager with 
proprietary software. Never attempt to dismantle the unit. There is a 
serious danger of a powerful electric shock. Never attempt to alter the 
software. 

2.

 

Do not leave or operate the unit in locations exposed to direct sunlight. 

3.

 

Do not store or use the unit close to a flame. 

4.

 

Do not drop the unit or otherwise subject it to strong impact. 

5.

 

QuickScan contains precision electronic components.  To ensure that 
scanning operations are performed correctly, do not subject the unit to 
impact or shock while a scan is being run. 

6.

 

Do not use or store the unit in very humid places; places subject to 
extreme temperature increases such as in direct sunlight, a closed car in 
summer, or extremely cold places; places subject to strong vibration; 
places affected by smoke or steam; places subject to strong magnetic 
fields (such as near motors, transformers, or magnets); sandy places, 
such as beaches or deserts, or places where there is wind-borne sand 
and dust. 

Summary of Contents for QuickSCan II

Page 1: ...USER MANUAL ACC 331 Doc M247 0318...

Page 2: ...ii EnviroLogix Inc 2018...

Page 3: ...8 Inventory 9 Help 10 Exit 10 Reading Test Strips 10 Reading Multi Matrix Barcode Cards 11 Interpreting Results 12 Results Table 12 Save Report 14 Print Report 15 Print Receipt 15 Export 15 Help 15 Cl...

Page 4: ...neous data storage where test results can be emailed printed or analyzed at any time A simple Microsoft Windows interface familiar to users worldwide and minimizing training time and A complete integr...

Page 5: ...by genetic modifications in crops mycotoxins algal toxins chemical contaminants and plant pathogens Interest in detecting and measuring these targets lies in many industries including agricultural bi...

Page 6: ...ion process contact the QuickScan Team at 1 866 408 4597 or techsupport envirologix com OPERATION Upon turning on the unit the QuickScan program will automatically open Or from the desktop double clic...

Page 7: ...eport button See Print Report below for more details To print a Report highlight a row from the desired report and click the Print button A dialog window will ask if you wish to print DataLog rows or...

Page 8: ...al output of your data depicting average results click the Analyze button Select test kit and starting and ending dates Select if you want running average and or supplier included on the graph Select...

Page 9: ...older or network location Currently it defaults to C EnviroLogix QuickScan Export The Results Table Column Titles setting allows the user to edit the column headers of entry fields Select the original...

Page 10: ...sers will see a daily reminder at 8 00 AM or upon startup thereafter to execute a Clean Test and Check Comb scan Changes to the time of the reminder can be made through Settings it is recommended that...

Page 11: ...y ID will autofill the Location field with the Program Key ID If checked the Location field will not be editable QSS On Screen Keyboard select to activate the QSS keyboard when entering data into user...

Page 12: ...s The cutoff will be applied to each individual result within the results screen even if multiple strips of the same test are scanned at one time Thus strips from multiple samples can be scanned at on...

Page 13: ...ntory tracking then click the Add button The kit will appear in the Test Kit column of the Inventory grid Enter the number of tubes canisters for single strip kits or number of pouches for combs into...

Page 14: ...by topic It also provides contact information for EnviroLogix Customer Service Each screen in the QuickScan program has a Help button When the Help button is clicked the program will open the file to...

Page 15: ...ncoded as barcodes on the Barcode Card included in the kit box The curves must be uploaded into the program prior to running the test strips To do this open the drawer and slide the MMBC face up into...

Page 16: ...he scale for mycotoxins is ppb or ppm For GMO tests results will fall between LOD the limit of detection and the upper limit of the quantitative range See the Product Insert for each test kit for more...

Page 17: ...938 The Results Table features five user input options intended to enhance the traceability of deliveries or shipments A drop down menu for the user entry columns will remember the last 30 entries mad...

Page 18: ...order in which strips are reported The QuickScan Results Screen provides a digital image of the scanned strip or comb The image shows the identification label barcode on the strip test lines and the s...

Page 19: ...ck the Export button to create an export file Export file type can be selected from the Export Format drop down located in the Settings tab Export files can be exported to a location defined using the...

Page 20: ...des with strip identifiers ID numbers unique to each device which are captured with the lot number within the DataLog When a product has been encoded with a Unique ID it is possible to clearly and qui...

Page 21: ...ld not be removed from the strips and should not be shared or exchanged across different strips Altering the barcode in any way will affect performance of the QuickScan System When processing strips b...

Page 22: ...date received if needed CLEANING AND SERVICING QuickScan II is designed to minimize dust or debris entering the scanner area and interfering with the reading of test strips It is important to keep th...

Page 23: ...urface place the White Card into the drawer and click the Clean Test button on the main screen A record of the Clean Test will be stored in the DataLog Scan the Check Comb Place the Check Comb into th...

Page 24: ...d in tray incorrectly Reposition strips and read again DIMENSIONS AND WEIGHT Length 37 cm Width 33 5 cm Height 20 cm Weight 11 15 lbs 5 06 kg SAFETY AND PRECAUTIONS 1 QuickScan is a precision system t...

Page 25: ...n writing of such defects during the warranty period and returns the unit to EnviroLogix for evaluation EnviroLogix will repair or replace the unit at its option During the period of defects evaluatio...

Reviews: