PORTABLE SYNGAS ANALYZER IR-GAS-600P
4
Safety information for Analyzer utilisation
IR-GAS-600P Analyser is not designed for use in hazardous area. Never use the analyzer in an
EX environment with potential gas explosion risks. Human safety could be seriously
endangered by flame, fire and/or deflagration.
IR-GAS-600P Analyser must be installed on a flat surface (if used as bench-type analyser) or
integrated in a gas analysis cabinet. Avoid locations where the analyzer can be exposed to
vibrations, chocks or intense electromagnetic fields as this can have a negative effect on the
performances of the analyzer.
Do not expose the analyzer to strong artificial light or sun radiations, wind, moisture of rain.
Prevent dust and moisture to enter the analyzer otherwise its good operation will be
impaired.
Gas connections and sample gas conditions
IR-GAS-600P SYNGAS analyser is usually supplied with our optional portable gas washing
unit
。
This unit can also be equipped with an optional powerful gas sampling pump
(requires 220 VAC power supply) to suit all process conditions. In case this accessory is
not supplied with the analyser, the customer shall ensure that the gas conditions at the
inlet of the analyser are in accordance with the following characteristics: free of solid
particles (<1μ), tar and of oil traces, with moisture content between 5 and 95% RH
non-condensing
Before switching ON the analyzer:
- Install the gas analyser on a flat surface
- Connect the gas discharge pipe on the gas out port of the analyser and position the
extremity of the pipe as far as possible from the analyser. Please also take care
about wind direction. Gas venting within an enclosed or non-ventilated area is not
allowed.
Switch ON the analyser
Wait for the complete cycle of preheating has elapsed, or 800 seconds.
Connect the outlet pipe of the gas washing unit to the input port of the analyser
Activate the internal pump of the gas analyser and adjust the flow meter of the gas
analyser at 1L/min. If the analyser is used with the gas washing unit with gas sampling
pump, the pump of the gas analyser does not need to be used.
If the gas is delivered in pressure (> 100 mbar relative) to the analyser, you shall place a
pressure limiting system before the analyser set to 100 mbar to protect it against
excessive pressure peaks (see conditions of the gas below). Excessive pressure may
cause leakage of toxic or explosive gases which can lead to serious accidents.
Sample gas conditions at GAS IN port of the analyser must answer to the following
specifications:
- Moisture: 5 to 95% RH non-condensing.
- Quality: free from solid particles, tar and oil traces
- Temperature: < 50°C