Safety information for Analyzer utilisation

IR-GAS-600P Analyser is not designed for use in hazardous area. Never use the analyzer in an
EX environment with potential gas explosion risks. Human safety could be seriously
endangered by flame, fire and/or deflagration.

IR-GAS-600P Analyser must be installed on a flat surface (if used as bench-type analyser) or
integrated in a gas analysis cabinet. Avoid locations where the analyzer can be exposed to
vibrations, chocks or intense electromagnetic fields as this can have a negative effect on the
performances of the analyzer.

Do not expose the analyzer to strong artificial light or sun radiations, wind, moisture of rain.

Prevent dust and moisture to enter the analyzer otherwise its good operation will be
impaired.

Gas connections and sample gas conditions

IR-GAS-600P SYNGAS analyser is usually supplied with our optional portable gas washing
unit

This unit can also be equipped with an optional powerful gas sampling pump

(requires 220 VAC power supply) to suit all process conditions. In case this accessory is
not supplied with the analyser, the customer shall ensure that the gas conditions at the
inlet of the analyser are in accordance with the following characteristics: free of solid
particles (<1μ), tar and of oil traces, with moisture content between 5 and 95% RH
non-condensing

Before switching ON the analyzer:
- Install the gas analyser on a flat surface
- Connect the gas discharge pipe on the gas out port of the analyser and position the

extremity of the pipe as far as possible from the analyser. Please also take care
about wind direction. Gas venting within an enclosed or non-ventilated area is not
allowed.

Switch ON the analyser

Wait for the complete cycle of preheating has elapsed, or 800 seconds.

Connect the outlet pipe of the gas washing unit to the input port of the analyser

Activate the internal pump of the gas analyser and adjust the flow meter of the gas
analyser at 1L/min. If the analyser is used with the gas washing unit with gas sampling
pump, the pump of the gas analyser does not need to be used.

If the gas is delivered in pressure (> 100 mbar relative) to the analyser, you shall place a
pressure limiting system before the analyser set to 100 mbar to protect it against
excessive pressure peaks (see conditions of the gas below). Excessive pressure may
cause leakage of toxic or explosive gases which can lead to serious accidents.

Sample gas conditions at GAS IN port of the analyser must answer to the following
specifications:

- Moisture: 5 to 95% RH non-condensing.
- Quality: free from solid particles, tar and oil traces
- Temperature: < 50°C

Summary of Contents for IR-GAS-600P

Page 1: ...IR GAS 600P USER MANUAL WUHAN ENVIRO SOLUTIONS TECHNOLOGY CO LTD Add The 4th floor No 28 Fozuling Road East lake Hi Tech Development Zone Industry Park Wuhan 430000 China Telephone 86 27 86639266 Mobi...

Page 2: ...e of the IR GAS 600P syngas portable analyser requires a full understanding of the analyzer operation Use the analyzer only as specified in this manual otherwise its good operation and initial perform...

Page 3: ...service attempts void this warranty Important The sample gas must be free from particles 1 oil traces and its moisture content if not WuHan Tenho can provide portable or on line gas conditioning syste...

Page 4: ...dance with the following characteristics free of solid particles 1 tar and of oil traces with moisture content between 5 and 95 RH non condensing Before switching ON the analyzer Install the gas analy...

Page 5: ...ly is switched OFF before connecting a data transmission line to the serial RS232 port Never use the analyser in an EX environment with potential gas explosion risks Notice for analyser application Sm...

Page 6: ...ways shut off the power supply before servicing the analyzer to prevent from electric shocks In order to operate correctly the analyzer must be maintained timely at least once a year Please consult yo...

Page 7: ...ical model is as follows most of the organic and inorganic multi atomic and molecular gases in the infrared region have a specific absorption wavelength When the infrared light passes through these ga...

Page 8: ...01 2 Note Measurement range can be customized by the requirement Max 6 components at the same time Response Time TD T90 10s NDIR Response Time 15min I O port RS232 Work temperature 0 50 Relative humid...

Page 9: ...PORTABLE SYNGAS ANALYZER IR GAS 600P 9...

Page 10: ...d be leaded to outdoor Gas outlet pipe should not no blocked 3 Sample gas pressure should be a little bigger than atmosphere pressure 4 Gas flow into the gas analyzer should control at 0 7L 1 2L min r...

Page 11: ...shows a few second after startup Sampling button press this button to open the pump Record button press this button to record the data The data will be recorded every 10 seconds Max 10 000 groups dat...

Page 12: ...S 600P 12 3 5 Measurement information In the INFO interface it shows gas sensor information The users cannot modify it only for reference 3 6 Calibration In the calibration interface It needs password...

Page 13: ...box 5 Function buttons to realize specific functions Zero when doing zero point calibration press zero button Span when doing span point calibration press span button Reset reset to factory calibratio...

Page 14: ...d as normal d Input the calibration gas value in the blank marked as 4 in the above picture and click SPAN button e g if CH4 is 25 calibration gas then input 25 e Then span calibration is finished 3 R...

Page 15: ...PORTABLE SYNGAS ANALYZER IR GAS 600P 15 3 7 1 USB connect computer Use USB cable into the USB interface the analyzer connected to the computer and then click on the Connect to PC via USB...

Page 16: ...sword Note please remember the modified password if you accidentally lost the password you can contact the company after service 3 7 3 Others 3 7 4 Delete data Click on Clear all data click the Confir...

Page 17: ...covery is successful indicating the successful recovery of factory settings 3 7 6 Touch screen calibration For touch screen calibration if the click speed is too fast may lead to calibration failure T...

Page 18: ...PORTABLE SYNGAS ANALYZER IR GAS 600P 18 3 7 7 Calorific value setting The settings are completed as shown in the following picture...

Page 19: ...flow less or no flow solutions Replacing dust filter paper Measurement value larger than the allowable error The instrument is not preheated enough Sampling instability Normal instruments start and p...

Page 20: ...he use it 3 The instrument should be stored in a clean ventilated and dry environment 4 Strong vibration should be prevented during transportation 6 Contact Information WUHAN ENVIRO SOLUTIONS TECHNOLO...

Page 21: ...PORTABLE SYNGAS ANALYZER IR GAS 600P 21...

Reviews: