Enviro IR-GAS-600P User Manual Download Page 2

PORTABLE SYNGAS ANALYZER IR-GAS-600P

2

1.

Foreword

1.1 Overview

Before using the analyzer and its accessories, please familiarize yourself with its operation by reading this
manual. If you have any questions, please do not hesitate to contact with us for assistance.

1.2 Unpacking Instructions

- Inspect the shipping boxes for any sign of damage. Damaged shipping boxes increase the risk to find

damaged material too.

- Remove the accessories as power cable and rack mounting lugs from the shipping box.
- Carefully remove the analyzer from its shipping box together with its protections.
- Remove the protections and place the material on a flat surface in a clean and dry location.
- Inspect the material for any sign of damage.

Note: keep the shipping box and the protective packaging for eventual future shipping needs.

1.3 Reporting Damage

If there are any apparent damages to the outside of the instrument due to shipping or handling, file a damage
claim and notify the forwarder immediately. The shipping container or packing materials should be retained for
inspection by the forwarder.

1.4 Before using the analyzer

The use of the IR-GAS-600P syngas portable analyser requires a full understanding of the analyzer operation.
Use the analyzer only as specified in this manual, otherwise its good operation and initial performances may
be impaired.

Authorized service representatives and original parts must only be employed in carrying out repairs to the
analyzer in order to maintain the validity of the warranty. Modification of components, use of non-original
parts, or use of incomplete or used parts will also invalidate the warranty.

Make sure there is no obstruction at the sample inlet before you operate the analyzer.

The sample gas must be free from particles (< 1µ), oil traces and its moisture content

Relative Humidity

must be between 5 and 95% non-condensing.

We highly recommend you to use the analyser exclusively

with our specific gas washing unit to clean the gas before it enters the analyzer

Use only original parts that have been specifically designed for your IR-GAS-600P portable syngas analyser.
The use of other parts than original ones may modify the performances of your analyzer.

We also recommend configuring the auto-zero function in order to make the zero drift negligible. Please
refer to the specific sections of this manual.

Do not expose the analyzer to electrical shock and/or severe continuous mechanical shock.

Summary of Contents for IR-GAS-600P

Page 1: ...IR GAS 600P USER MANUAL WUHAN ENVIRO SOLUTIONS TECHNOLOGY CO LTD Add The 4th floor No 28 Fozuling Road East lake Hi Tech Development Zone Industry Park Wuhan 430000 China Telephone 86 27 86639266 Mobi...

Page 2: ...e of the IR GAS 600P syngas portable analyser requires a full understanding of the analyzer operation Use the analyzer only as specified in this manual otherwise its good operation and initial perform...

Page 3: ...service attempts void this warranty Important The sample gas must be free from particles 1 oil traces and its moisture content if not WuHan Tenho can provide portable or on line gas conditioning syste...

Page 4: ...dance with the following characteristics free of solid particles 1 tar and of oil traces with moisture content between 5 and 95 RH non condensing Before switching ON the analyzer Install the gas analy...

Page 5: ...ly is switched OFF before connecting a data transmission line to the serial RS232 port Never use the analyser in an EX environment with potential gas explosion risks Notice for analyser application Sm...

Page 6: ...ways shut off the power supply before servicing the analyzer to prevent from electric shocks In order to operate correctly the analyzer must be maintained timely at least once a year Please consult yo...

Page 7: ...ical model is as follows most of the organic and inorganic multi atomic and molecular gases in the infrared region have a specific absorption wavelength When the infrared light passes through these ga...

Page 8: ...01 2 Note Measurement range can be customized by the requirement Max 6 components at the same time Response Time TD T90 10s NDIR Response Time 15min I O port RS232 Work temperature 0 50 Relative humid...

Page 9: ...PORTABLE SYNGAS ANALYZER IR GAS 600P 9...

Page 10: ...d be leaded to outdoor Gas outlet pipe should not no blocked 3 Sample gas pressure should be a little bigger than atmosphere pressure 4 Gas flow into the gas analyzer should control at 0 7L 1 2L min r...

Page 11: ...shows a few second after startup Sampling button press this button to open the pump Record button press this button to record the data The data will be recorded every 10 seconds Max 10 000 groups dat...

Page 12: ...S 600P 12 3 5 Measurement information In the INFO interface it shows gas sensor information The users cannot modify it only for reference 3 6 Calibration In the calibration interface It needs password...

Page 13: ...box 5 Function buttons to realize specific functions Zero when doing zero point calibration press zero button Span when doing span point calibration press span button Reset reset to factory calibratio...

Page 14: ...d as normal d Input the calibration gas value in the blank marked as 4 in the above picture and click SPAN button e g if CH4 is 25 calibration gas then input 25 e Then span calibration is finished 3 R...

Page 15: ...PORTABLE SYNGAS ANALYZER IR GAS 600P 15 3 7 1 USB connect computer Use USB cable into the USB interface the analyzer connected to the computer and then click on the Connect to PC via USB...

Page 16: ...sword Note please remember the modified password if you accidentally lost the password you can contact the company after service 3 7 3 Others 3 7 4 Delete data Click on Clear all data click the Confir...

Page 17: ...covery is successful indicating the successful recovery of factory settings 3 7 6 Touch screen calibration For touch screen calibration if the click speed is too fast may lead to calibration failure T...

Page 18: ...PORTABLE SYNGAS ANALYZER IR GAS 600P 18 3 7 7 Calorific value setting The settings are completed as shown in the following picture...

Page 19: ...flow less or no flow solutions Replacing dust filter paper Measurement value larger than the allowable error The instrument is not preheated enough Sampling instability Normal instruments start and p...

Page 20: ...he use it 3 The instrument should be stored in a clean ventilated and dry environment 4 Strong vibration should be prevented during transportation 6 Contact Information WUHAN ENVIRO SOLUTIONS TECHNOLO...

Page 21: ...PORTABLE SYNGAS ANALYZER IR GAS 600P 21...

Reviews: