Envirco Hospi-Gard Operation Manual Download Page 15

H

OSPI

-G

ARD

®

 I

SO

C

LEAN

®

 

and 

I

SO

C

LEAN

® 

UV

Operation Manual

Page 11

CAUTION:  Proper protective equipment and measures must be used at all times

during unit cleaning and maintenance.  Check with your Safety Office to

assure the cleaning solutions, protective equipment (disposable hospital

gown, HEPA respirator, protective gloves) and protocols followed comply

with your facility’s and CDC guidelines.

CAUTION:

Disconnect the unit from the electrical power source before attempting

service. Do not work inside the IsoClean

®

 UV with the UV light on. Proper

replacement of UV containment baffles, after maintenance, is mandatory to

maintain safe short-term and long-term UV exposure levels.

4.2.  Cleaning the Unit

4.2.1.  Periodic cleaning of the unit may be done with a mild disinfecting soap and water

solution.  A 1% or 1/100 dilution of bleach is an optional disinfectant

5

.

4.2.2.  Cleaning must be performed with the unit running.  Wipe the exterior of the unit

with a damp sponge or cloth, paying particular attention to the prefilter intake
grille.

4.3.  Prefilter Replacement

4.3.1.  The prefilter should be replaced every 60 to 90 days.  Regularly replacing the

prefilter will maximize the life of the HEPA filter.

The IsoClean

®

 UV

 

comes standard with an anti-microbial pleated prefilter. It is

recommended that a pleated prefilter is used to insure that safe long-term
recommended exposure levels to UV are maintained.

4.3.2.  Wipe the unit as described in Section 4.2.

4.3.3.  Spray a mist of phenolic disinfectant such as “Amplyl” or “Lysol” into the air

intake grille.

4.3.4.  Remove the two (2) thumb nuts from the face of the intake grille and pull the

grille out.

4.3.5.  Insert gloved hand into a red plastic, medical waste disposal bag.  Grasp the top

portion of the prefilter with bagged hand and gently pull outward. The airflow of
the unit will offer a resistance to pulling the prefilter out.

4.3.6.  Turn the bag inside out over the prefilter then seal the bag for containment.

4.3.7.  Install a new prefilter and close the grille up in the reverse order of removal.

__________________________________________________________________________________

5

 

Rutala WA, APIC guideline for selection and use of disinfectants, Am J Infect Control 1996 Aug ; 24(4): 313-42.

Summary of Contents for Hospi-Gard

Page 1: ...www envirco com Operation Manual HOSPI GARD IsoClean And IsoClean UV...

Page 2: ...vice panel and lock service panel to prevent power from being switched on accidentally and follow proper procedures as necessary E Use this unit only in the manner intended by the manufacturer If you...

Page 3: ...t Mode Negative Pressure 6 3 7 Negative Pressure 7 4 0 Maintenance and Service 9 4 1 Parts Identification 9 4 2 Cleaning the Unit 11 4 3 Prefilter Replacement 11 4 4 Removal of Front and Rear Panels 1...

Page 4: ...ing Diagram 19 6 1 Hospi Gard IsoClean Wiring Diagram 19 6 2 Hospi Gard IsoClean UV Wiring Diagram 20 7 0 Specifications 21 7 1 IsoClean 21 7 2 IsoClean UV 21 8 0 Replacement Parts 22 9 0 Airflow Meas...

Page 5: ...CFM on low 700 CFM on medium and up to 800 CFM on high to carry away airborne particulate generated by activities within the room The IsoClean UV offers the added benefits of ultraviolet irradiation o...

Page 6: ...solation Rooms Waiting Rooms Bronchoscopy Rooms Emergency Rooms Intensive Care Units Sputum Induction Booths and Rooms Aerosol Pentamidine Booths and Rooms Renal Dialysis Treatment Rooms Hospital Reco...

Page 7: ...material received should be compared and any shortages reported to Envirco immediately 2 2 Assembly 2 2 1 The Hospi Gard IsoClean is shipped completely assembled from the factory and requires no asse...

Page 8: ...an unit is set on low at the factory 3 2 1 To change speeds remove the exhaust grille by pulling down and lifting out 3 2 2 Turn the 3 speed switch to the desired setting 3 2 3 Replace the exhaust gri...

Page 9: ...ospi Gard IsoClean speed setting and the size of the room See Table 1 in the Appendix for representative room air changes per hour ACH 3 4 3 For maximum recirculation efficiency the unit should be pla...

Page 10: ...3 5 6 The IsoClean is designed to exhaust 100 300 CFM at the 6 inch exhaust port It is recommended that the unit be set on the high speed setting when operating in this mode to achieve maximum exhaust...

Page 11: ...egative pressure relative to the corridor or other areas connected to the room The minimum pressure difference necessary to achieve and maintain negative pressure that will result in airflow into the...

Page 12: ...he responsibility of the purchaser and not Envirco Corporation to ensure the room in which the unit is used is in compliance with CDC Guidelines4 for a negative pressure isolation room and monitored t...

Page 13: ...PI GARD ISOCLEAN and ISOCLEAN UV Operation Manual Page 9 MAINTENANCE AND SERVICE 4 0 Maintenance and Service 4 1 Parts Identification 4 1 1 Hospi Gard IsoClean Figure 2 Hospi Gard IsoClean Parts Locat...

Page 14: ...HOSPI GARD ISOCLEAN and ISOCLEAN UV Operation Manual Page 10 4 1 2 Hospi Gard IsoClean UV Figure 3 Hospi Gard IsoClean UV Parts Location...

Page 15: ...r intake grille 4 3 Prefilter Replacement 4 3 1 The prefilter should be replaced every 60 to 90 days Regularly replacing the prefilter will maximize the life of the HEPA filter The IsoClean UV comes s...

Page 16: ...rs depending on the environment and prefilter maintenance schedule The filter indicator light will give a warning of filter loading by flickering on high speed When the light stays lit on high speed t...

Page 17: ...strain relief on the side of the unit 4 6 4 Remove the old cord and replace with the new one following the steps above in reverse order 4 7 On Off Switch 4 7 1 With the back panel removed unplug the...

Page 18: ...ve the wheel 4 12 3 Loosen the screw on the motor belly band until the motor can slide out 4 12 4 Replace with the new motor following the steps above in reverse order ensuring there is an inch to inc...

Page 19: ...15 1 3 Multiply the total volume of the unit by 0 3g ft3 to determine the gram weight of Paraformaldehyde required Decontamination of the IsoClean requires 3 9 grams 12 ft3 x 0 3g ft3 of Paraformaldeh...

Page 20: ...mprove ventilation 4 15 1 15 Allow the unit to ventilate overnight 4 15 1 16 Remove the unit from exhaust ventilation when the formaldehyde gas has been exhausted CAUTION During unit decontamination r...

Page 21: ...HEPA filter and grasp the old light tube gently Rotate the light tube 90 degrees in either direction to enable removal you should feel tension on the tube release To remove the light tube carefully p...

Page 22: ...tion 6 Replace or repair faulty component s 5 1 4 Check the motor and replace if necessary 5 1 5 Check the motor capacitor and replace if necessary 5 2 Low Air Flow 5 2 1 Check prefilter for obstructi...

Page 23: ...HOSPI GARD ISOCLEAN and ISOCLEAN UV Operation Manual Page 19 WIRING DIAGRAM 6 0 Wiring Diagram 6 1 Hospi Gard IsoClean Wiring Diagram Figure 4 Hospi Gard IsoClean Wiring Diagram...

Page 24: ...HOSPI GARD ISOCLEAN and ISOCLEAN UV Operation Manual Page 20 6 2 Hospi Gard IsoClean UV Wiring Diagram Figure 5 Hospi Gard IsoClean UV Wiring Diagram...

Page 25: ...m efficiency Prefilter Woven polyester with frame Motor 1 3 Hp direct drive three speed continuous duty Exhaust Port 6 Inch Duct Collar 7 2 IsoClean UV 10850 005 Dimensions 62 25 H x 23 75 W x 19 5 D...

Page 26: ...No Optional Accessories Quantity 11055 8 in A C Collar with Flange for 100 Exhaust IsoClean 1 63338 Annunciator 1 63352 Antimicrobial Prefilter 1 63349 Carbon Prefilter 1 63337 Room Pressure Monitor 1...

Page 27: ...an UV 1 24392 001 UV Containment Baffle for IsoClean UV 1 Part No Optional Accessories Quantity 23877 002P 8 inch Duct Collar Kit for IsoClean UV 1 63338 Annunciator 1 63337 Room Pressure Monitor 1 63...

Page 28: ...HOSPI GARD ISOCLEAN and ISOCLEAN UV Operation Manual Page 24 TESTING 9 0 Airflow Measurement and Testing...

Page 29: ...t 3 12ftx15ftx8ft 1440 ft 3 12ftx15ftx9ft 1620 ft 3 15ftx15ftx8ft 1800 ft 3 15ftx20ftx8ft 2400 ft 3 20ftx20ftx8ft 3200 ft 3 Low 550 CFM 41 ACH 28 ACH 23 ACH 20 ACH 18 ACH 13 ACH 10 ACH Medium 700 CFM...

Page 30: ...d to the cost of transporting the defective equipment to Envirco s repair facility The buyer s sole remedy under this limited warranty is the repair or replacement of any defective part of the equipme...

Reviews: