Hospi-Gard
®
IsoClean
®
800 & 400
HEPA Filtration System
8
Innovators in Clean Air Technology | www.envirco.com
I n s t a l l a t i o n , O p e r a t i o n , & M a i n t e n a n c e M a n u a l
CAUTION: PROPER PROTECTIVE EQUIPMENT AND MEASURES MUST BE USED AT ALL TIMES DURING UNIT
CLEANING AND MAINTENANCE. CHECK WITH YOUR SAFETY OFFICE TO ASSURE THE CLEANING SOLUTIONS,
PROTECTIVE EQUIPMENT (DISPOSABLE HOSPITAL GOWN, HEPA RESPIRATOR, PROTECTIVE GLOVES) AND
PROTOCOLS FOLLOWED COMPLY WITH YOUR FACILITY’S AND CDC GUIDELINES.
CAUTION: DISCONNECT THE UNIT FROM THE ELECTRICAL POWER SOURCE BEFORE ATTEMPTING SERVICE. DO NOT
WORK INSIDE THE ISOCLEAN UV WITH THE UV LIGHT ON. PROPER REPLACEMENT OF UV CONTAINMENT BAFFLES
AFTER MAINTENANCE IS MANDATORY TO MAINTAIN SAFE SHORT-TERM AND LONG-TERM UV EXPOSURE LEVELS.
Cleaning the Unit
1.
Periodic Cleaning of the unit may be done with a mild disinfecting soap and water solution. A 1% or 1/100 dilution of bleach
is an optional disinfectant
3
.
2.
Exterior cabinet cleaning must be performed with the unit running. Wipe the exterior of the unit with a damp sponge or
cloth, paying particular attention to the prefilter intake grille.
3
Rutala WA, APlC guideline for selection and use of disinfectants, Am J lnfect Control 1996 Aug ; 24(4): 313-42.