Tab 2: Calibration
(Differences between the field housing and software now result here.)
2.1
Calibration factor
Input:
0.01 … 99.99
Adjustment of the measurement value,
multiplies the (Field housing and software)
calibrated measurement value, serves for
subsequent correction of the calibration.
2.2
Calibration filter [s]
Input:
1 … 9999
Quiescent filter for the calibration log.
2.3
Calibration points
Input:
2 … 5
Selection of the calibration points in the
calibration table.
2.4 Calibration
Calibration submenu
2.4.1 P1 value
Input:
Output measurement value in the selected
measurement value
mass / time unit.
to be shown
(2.4.2) P1 calibration
Adoption:
Raw value
Adoption of the instantaneous raw value
current raw value
(filtered) from the mass flow with the key
[
←
]. The value can also be entered
manually.
. . . (depending on the number of support points)
For additional support points (depending on
[ 2.3 ]), additional value pairs can be set.