MN-38787  •  Rev 0  •  05/17  •  WarmWatch® Temperature Reader

Operation

How to set the Injection Mode temperature

Press the On/Off button. The LED screen 

illuminates.
Press the Injection button to select the Injection 

Fluids Mode. The Injection Fluids Mode Indicator 

illuminates.
Press the Up Arrow button or Down Arrow 

button to adjust the injection temperature as 

needed.
Allow the fl uid chamber to reach the set 

temperature.

How to display the temperature reading

Press the Temperature Recall button to view the 

actual temperature captured with the chamber 

sensor. The actual chamber temperature is 

displayed for fi ve (5) seconds before displaying 

the fl uid chamber temperature set-point.

Place a warm injection fl uid bag on the 

WarmWatch Temperature Reader. Close the fl uid 

chamber door. Press and hold the Temperature 

Recall

 button for three (3) seconds to display 

the WarmWatch® temperature reading of the 

injection fl uid bag. The WarmWatch temperature 

reading is displayed twenty (20) seconds before 

returning to the fl uid chamber temperature 

set-point.

How to Calibrate the WarmWatch® Temperature 

Reader

 

NOTE

:   

The fl uid chamber must be calibrated prior 

to the WarmWatch Temperature Reader.

1.  Place the temperature reader in the fl uid 

chamber and attach the cord to the Ethernet port 

in the back of the chamber.

2.  Press the On/Off button. The LED screen 

illuminates. 

3.  Press the Injection button to select the Injection 

Fluids Mode. The Injection Fluids Mode Indicator 

illuminates. 

4.  Press the Up Arrow button or Down Arrow 

button to adjust the injection temperature to 

40°C (104°F).

5.  When the fl uid chamber reaches 40°C (104°F), 

press and hold the Temperature Recall 

button for three (3) seconds. The WarmWatch 

temperature reading is displayed.

6.  While the WarmWatch temperature reading is 

displayed, press and hold both the Up Arrow and 

Down Arrow

 buttons until CALb appears in the 

LED screen.

7.  Release the Up Arrow and Down Arrow 

buttons. The WarmWatch Temperature Reader 

self-calibrates. The fl uid chamber set injection 

temperature is displayed.

Up Arrow

Button

Down Arrow

Button

On/Off

Button

Button Lockout

Indicator Light

Error

Indicator Light

Electronic Control and LED Display

Temperature

Recall Button

LED 

Screen

Mode

Indicators

Irrigation 

Button

Injection 

Button

Temperature Range:

 

•  Injection (INJ) temperature range: 

  32°C to 40°C (90°F to 104°F)

Summary of Contents for WarmWatch

Page 1: ...fluid to patients WarmWatch is fully integrated into the fluid warmer control Temperature range 32 C 40 C 90 F 104 F View the individual bag surface temperature while the bag is inside the fluid warming compartment The temperature is displayed within seconds after the touch of a button on the fluid warmer control Curved lightweight base cradles bags up to 3 liters in size ...

Page 2: ...ature Reader Table of Contents Delivery 3 Transportation Damage and Claims 3 Unpacking 4 Installation 5 Operation Instructions 6 Cleaning 7 Limited Warranty 8 Authorized Representative MDSS GmbH Schiffgraben 41 30175 Hannover Germany REP EC ...

Page 3: ...e occur in shipment it is a matter between the carrier and the consignee In such cases the carrier is assumed to be responsible for the safe delivery of the merchandise unless negligence can be established on the part of the shipper 1 Conduct an immediate inspection immediately after it is moved to the receiving area Do not wait until after the appliance or accessory is moved to a storage area 2 D...

Page 4: ...mage personal injury and may void the warranty 3 Do not discard this manual This manual is considered part of the accessory and is to be provided to the owner or manager of the business or to the person responsible for training operators Additional manuals are available from the service department 4 Remove all protective plastic film and packaging materials from the accessory before installing it ...

Page 5: ...erature reader with a clean damp cloth to remove all detergent residue Dry the temperature reader with a clean lint free cloth 4 Open the fluid chamber door 5 Open the Ethernet port cover located at the rear of the fluid chamber 6 Insert one end of the cord connector into the Ethernet port at the rear of the fluid chamber 7 Insert the opposite end of the cord connector into the port at the rear of...

Page 6: ...WarmWatch Temperature Reader NOTE The fluid chamber must be calibrated prior to the WarmWatch Temperature Reader 1 Place the temperature reader in the fluid chamber and attach the cord to the Ethernet port in the back of the chamber 2 Press the On Off button The LED screen illuminates 3 Press the Injection button to select the Injection Fluids Mode The Injection Fluids Mode Indicator illuminates 4...

Page 7: ...void using cleaning materials with sharp edges How to Clean the WarmWatch Temperature Reader 1 Remove the detachable coil cord from the WarmWatch base 2 Wash the coil cord with hot soapy water Apply the hot soapy water with a clean damp cloth Wipe with a clean damp cloth to remove all detergent residue Dry the coil cord with a clean lint free cloth 3 Clean the exterior of the temperature reader wi...

Page 8: ...or quaternary salts poor water quality or equipment with missing or altered serial numbers 4 Any losses or damage resulting from malfunction including loss of contents or consequential or incidental damages of any kind 5 Equipment modified in any manner from original model substitution of parts other than factory authorized parts removal of any parts including legs or addition of any parts 6 Colla...