background image

G E N E R A L I N F O R M A T I O N

4

ENTHERMICS MEDICAL SYSTEMS

O C M / E C 1 5 4 0

bl

0 1 / 0 7

This warming cabinet is designed to safely store warm

either irrigation fluids or injection fluids and blankets.

The dual-chambered warming cabinet is constructed

with a 20 gauge stainless steel exterior casing and door with

handles and hinges designed to withstand heavy usage. A

door with window allows observation of inventory with the

door closed. The cabinet is warmed using the patented Halo

Heat

®

low-heat-density electrothermal cable array. The

electrothermal cable is positioned in the floor and two sides

of the warming cabinet, providing even heating of the

interior chamber. Each chamber temperature is regulated by

a

separate

electronic control. A fan located inside the

chamber

mixes

the air to prevent temperature stratification

and to ensure an accurate chamber temperature within

±2°F (±1.1°C) of the set point for temperatures up to

110°F (43°C) and within ±3°F (±1.6°C) of the set point for

temperatures above 110° to 150°F (43°C to 66°C). Each

chamber has an individual electronic control consisting

of a 4 digit L.E.D. display, ON/OFF key, INCREASE and

DECREASE keys, integrated LOCK feature and a series of

prompt sequence indicators.

The (upper) blanket warming chamber has an adjustable

temperature range of 90° to 200°F (32° to 94°C). This

chamber also has a TIMER mode that allows the user to

program the control to automatically turn on and turn off

once during a 24-hour period at a selected time. This

enables the blanket warming chamber (

only

) to be shut off

automatically at night to save energy, but to turn on again in

the early morning to ensure warm blankets are available.

The (lower) fluid warming chamber can be programmed

to warm

either

irrigation fluids (IRR)

or

injection fluids

(INJ), with separate temperature ranges provided depending

on the choice selected. IRR temperature may be adjusted

from 90° to 150°F (32° to 66°C), and the INJ temperature

can be adjusted from 90° to 110°F (32° to 43°C). An alarm

will sound if temperatures exceed 10°F (6°C) over the set-

point temperature, and a TEMP indicator will blink

indicating an over-temperature condition.

A warming shut-off system, separate from the electronic

controls, prevents overheating. The electronic controls can

easily be set to operate in Fahrenheit or Celsius. After a

power failure, the cabinet will remember its programming

and begin to operate as before. The ON/OFF indicator will

blink to indicate a failure occurred; pressing the ON/OFF

key once will eliminate this blinking.

The blanket warming chamber contains an epoxy-coated

blanket support assembly and one (1) shelf. The fluid

warming chamber is equipped with three (3) white, epoxy-

coated wire baskets to accommodate fluids packaged in bags

or bottles, mounted on basket rail supports. Each basket has

a

14 liter maximum capacity

. The cabinet is furnished

with a full perimeter rubber bumper assembly for exterior

protection, and one (1) set of 5" (127mm) heavy-duty

casters, two with locking brake.

AT NO TIME SHOULD THE INTERIOR OR 
EXTERIOR BE STEAM CLEANED, HOSED 
DOWN, OR FLOODED WITH WATER OR 
LIQUID SOLUTION OF ANY KIND.  DO NOT 
USE WATER JET TO CLEAN.

S E V E R E   D A M A G E   O R   E L E C T R I C A L   H A Z A R D  

C O U L D   R E S U LT.

 

WARRANTY BECOMES VOID IF APPLIANCE IS FLOODED.

THIS UNIT HAS NOT BEEN 

APPROVED FOR WARMING OF 

BLOOD OR BLOOD PRODUCTS.

Summary of Contents for EC1540bl

Page 1: ... Water Street P O Box 443 Menomonee Falls Wisconsin 53052 0443 U S A P H O N E 2 6 2 2 5 1 8 3 5 6 8 0 0 T O B W A R M U S A CANADA F A X 2 6 2 2 5 1 7 0 6 7 w w w e n t h e r m i c s c o m 8 0 0 8 6 2 9 2 7 6 U S A CANADA 8 0 0 3 2 9 8 7 4 4 Combination Blanket Fluid Warming Cabinet EC1540bl 120V 230V ...

Page 2: ...ranty card to the factory as soon as possible to assure prompt service in the event of a warranty parts and labor claim This manual must be read and understood by all people using or installing the equipment model Contact the Enthermics service department if you have any questions concerning installation operation or maintenance NOTE All claims for warranty must include the full model number and s...

Page 3: ...n blankets ONLY No other use for this device is authorized or recommended Enthermics fluid warmers are ONLY intended for warming medical solutions for irrigation and injection The IRRIGATION MODE should be selected for warming irrigation fluids and the INJECTION MODE should be selected for warming injection fluids Please refer to the labeling of the manufacturer of the products to be warmed regard...

Page 4: ... air or with oxygen or nitrous oxide THIS COULD RISK AN EXPLOSION INJECTION FLUIDS SHOULD NOT BE WARMED OVER 110 F 43 C THIS UNIT HAS NOT BEEN APPROVED FOR WARMING OF BLOOD OR BLOOD PRODUCTS To prevent an electrical shock hazard between the appliance and other appliances or metal parts in close vicinity an equalization bonding stud is provided An equalization bonding lead must be connected to this...

Page 5: ...tically at night to save energy but to turn on again in the early morning to ensure warm blankets are available The lower fluid warming chamber can be programmed to warm either irrigation fluids IRR or injection fluids INJ with separate temperature ranges provided depending on the choice selected IRR temperature may be adjusted from 90 to 150 F 32 to 66 C and the INJ temperature can be adjusted fr...

Page 6: ... 800 730 81mm 25 100 636 40mm 30 000 463 04mm 11 000 280 03mm 2 200 56 34mm 18 900 480 29mm Cavity 22 125 561 98mm 67 000 1702 61mm 5 500 139 22mm 25 200 639 22mm 26 900 683 49mm Cavity 26 900 683 49mm Cavity 72 500 1841 83mm ON OFF OVER TEMP IRR INJ LOCK OVERTEMP ERROR ON OFF LOCK TIME START STOP TIMER 11 000 280 03mm ...

Page 7: ...tatus indicator will be flashing B Display will indicate 128 or another number alternating with the setting Please see important note below if a number other than 128 is displayed C Control will beep every third flash of the number displayed Press the ON OFF key once to acknowledge that the power has been restored The ON OFF status indicator will stop flashing and the beep will be silenced The dis...

Page 8: ...ndicator will illuminate Attempts to operate the ON OFF key or to change the temperature set point will be unsuccessful To unlock the warmer control press and hold the ON OFF key and the DOWN arrow button at the same time The unit will beep the control will unlock and the LOCK indicator will extinguish ON OFF LOCK TIME START STOP L E D DISPLAY STATUS INDICATOR L E D UP ARROW DOWN ARROW AUTOTIMER K...

Page 9: ...the set point temperature OVERTEMP When the control senses a temperature 10 F 6 C greater than the set point this indicator will illuminate An alarm will also sound The alarm can be muted by pressing any key This indicator will extinguish and the green indicator of the overtemp key will stop blinking once the warmer temperature drops back to the set point temperature POWER FAIL DETECT If the power...

Page 10: ...Exercise judgment to determine inventory rotation protocols and warm up time for the fluids you use Caution Check fluid temperature prior to use If the warmer control has failed or if error messages are displayed discard your inventory and contact your service representative F L U I D O P E R A T I O N A L P R O C E D U R E S L O W E R C H A M B E R POWER CIRCUIT BREAKER SWITCH REFER TO FLUID MANU...

Page 11: ...earance and life of this appliance Spilled items should be removed and the area wiped as soon as possible but at the very least a minimum of once a day Always thoroughly rinse surfaces after using a cleaning agent and wipe standing water as quickly as possible after rinsing CLEANING AGENTS Use non abrasive cleaning products designed for use on stainless steel surfaces Cleaning agents must be chlor...

Page 12: ...Interior can be wiped with a sanitizing solution after cleaning and rinsing This solution must be approved for use on stainless steel surfaces Replace blanket support assembly 5 To help maintain the protective film coating on polished stainless steel clean the exterior of the cabinet with a cleaner recommended for stainless steel surfaces Spray the cleaning agent on a clean cloth and wipe with the...

Page 13: ...m meter to measure the resistance of the sensor Check sensor at 32 F 0 C using a container of ice water If Ohm reading is 100 replace display If Ohm reading is 100 replace sensor Check wires for integrity Check for proper and secure connections at the thermostat and terminal block If necessary re secure the faulty connections If error continues call Service E 31 Over Temp Unit may be overloaded Re...

Page 14: ... E1131 28 STRAIN RELIEF BUSHING 1 BU 3964 29 TERMINAL BLOCK 3 FORM COMPRESS 1 BK 3019 30 TERMINAL BLOCK PORCELAIN 2 BK 33546 31 THERMOSTAT HI LIMIT FLUID 120V 2 E3040TT 32 THERMOSTAT HI LIMIT BLANKET 120V 2 TT 34350 33 THERMOSTAT HI LIMIT FLUID 230V 1 E3040TT 34 THERMOSTAT HI LIMIT BLANKET 230V 1 TT 34058 35 THERMOSTAT MANUAL RESET 230V 2 TT 33992 DESCRIPTION QTY PART NO MECHANICAL HARDWARE 36 BLA...

Page 15: ...L WITH BRAKE GI 25942 DRAWER SLIDE GUIDE SH 25908 BLANKET SUPPORT ASSEMBLY 5001253 WINDOW DOOR ASSEMBLY RH GS 23380 DOOR GASKET ASSEMBLY UPPER SN 33540 SENSOR 5002972 BASKET SUPPORT ASSEMBLY SH 25911 BLANKET SHELF HG 22338 HINGES 5004206 LAMP GUARD ASSEMBLY HG 22338 HINGES BS 26690 BASKETS LP 33267 LAMP ASSEMBLY 5004206 LAMP GUARD ASSEMBLY GD 2396 WIRE GUARD ONLY SN 33541 SENSOR 5001776 FULL PERIM...

Page 16: ...MINAL BLOCK E3038RL RELAY 12V DC COIL BP 3567 BEEPER RL 33829 RELAY 230V BA 34041 POWER SUPPLY BOARD E3038RL RELAY 12V DC COIL E3029CD CORDSET HOSPITAL GRADE BU 3964 STRAIN RELIEF BUSHING CR 33717 CONNECTOR 10 PIN CR 33763 CONNECTOR 4 PIN CR 33718 CONNECTOR 9 PIN CC 34423 CONTROL ASSEMBLY Service View of EC1540bl 120v ...

Page 17: ...OCM EC1540bl 01 07 Service View of EC1540bl 120v CS 24874 CASTERS 5 127mm RIGID 5003418 CORD GRIP ASSEMBLY FA 3973 BOX FAN 5001776 FULL PERIMETER RUBBER BUMPER ASSEMBLY SW 33858 CIRCUIT BREAKER SWITCH E3029CD CORDSET HOSPITAL GRADE ...

Page 18: ...RL RELAY 12V DC COIL BP 3567 BEEPER RL 33829 RELAY 230V BA 34041 POWER SUPPLY BOARD E3038RL RELAY 12V DC COIL E3029CD CORDSET HOSPITAL GRADE BU 3964 STRAIN RELIEF BUSHING CR 33717 CONNECTOR 10 PIN CR 33763 CONNECTOR 4 PIN CR 33718 CONNECTOR 9 PIN CC 34423 CONTROL ASSEMBLY Service View of EC1540bl 230v TT 33992 THERMOSTAT MANUAL RESET ...

Page 19: ...07 Service View of EC1540bl 230v CS 24874 CASTERS 5 127mm RIGID 5003418 CORD GRIP ASSEMBLY FA 3974 BOX FAN 5001776 FULL PERIMETER RUBBER BUMPER ASSEMBLY SW 33858 CIRCUIT BREAKER SWITCH E3029CD CORDSET HOSPITAL GRADE TT 33992 THERMOSTAT MANUAL RESET ...

Page 20: ...19 ENTHERMICS MEDICAL SYSTEMS OCM EC1540bl 01 07 ...

Page 21: ...20 ENTHERMICS MEDICAL SYSTEMS OCM EC1540bl 01 07 ...

Page 22: ...first This warranty does not apply to 1 Calibration 2 Equipment damage caused by accident shipping improper installation or alteration 3 Equipment used under conditions of abuse misuse carelessness or abnormal conditions including equipment subjected to harsh or inappropriate chemicals including but not limited to compounds containing chloride or quaternary salts poor water quality or equipment wi...

Reviews: