G E N E R A L I N F O R M A T I O N
4
ENTHERMICS MEDICAL SYSTEMS
O C M / E C 1 5 4 0
bl
0 1 / 0 7
This warming cabinet is designed to safely store warm
either irrigation fluids or injection fluids and blankets.
The dual-chambered warming cabinet is constructed
with a 20 gauge stainless steel exterior casing and door with
handles and hinges designed to withstand heavy usage. A
door with window allows observation of inventory with the
door closed. The cabinet is warmed using the patented Halo
Heat
®
low-heat-density electrothermal cable array. The
electrothermal cable is positioned in the floor and two sides
of the warming cabinet, providing even heating of the
interior chamber. Each chamber temperature is regulated by
a
separate
electronic control. A fan located inside the
chamber
mixes
the air to prevent temperature stratification
and to ensure an accurate chamber temperature within
±2°F (±1.1°C) of the set point for temperatures up to
110°F (43°C) and within ±3°F (±1.6°C) of the set point for
temperatures above 110° to 150°F (43°C to 66°C). Each
chamber has an individual electronic control consisting
of a 4 digit L.E.D. display, ON/OFF key, INCREASE and
DECREASE keys, integrated LOCK feature and a series of
prompt sequence indicators.
The (upper) blanket warming chamber has an adjustable
temperature range of 90° to 200°F (32° to 94°C). This
chamber also has a TIMER mode that allows the user to
program the control to automatically turn on and turn off
once during a 24-hour period at a selected time. This
enables the blanket warming chamber (
only
) to be shut off
automatically at night to save energy, but to turn on again in
the early morning to ensure warm blankets are available.
The (lower) fluid warming chamber can be programmed
to warm
either
irrigation fluids (IRR)
or
injection fluids
(INJ), with separate temperature ranges provided depending
on the choice selected. IRR temperature may be adjusted
from 90° to 150°F (32° to 66°C), and the INJ temperature
can be adjusted from 90° to 110°F (32° to 43°C). An alarm
will sound if temperatures exceed 10°F (6°C) over the set-
point temperature, and a TEMP indicator will blink
indicating an over-temperature condition.
A warming shut-off system, separate from the electronic
controls, prevents overheating. The electronic controls can
easily be set to operate in Fahrenheit or Celsius. After a
power failure, the cabinet will remember its programming
and begin to operate as before. The ON/OFF indicator will
blink to indicate a failure occurred; pressing the ON/OFF
key once will eliminate this blinking.
The blanket warming chamber contains an epoxy-coated
blanket support assembly and one (1) shelf. The fluid
warming chamber is equipped with three (3) white, epoxy-
coated wire baskets to accommodate fluids packaged in bags
or bottles, mounted on basket rail supports. Each basket has
a
14 liter maximum capacity
. The cabinet is furnished
with a full perimeter rubber bumper assembly for exterior
protection, and one (1) set of 5" (127mm) heavy-duty
casters, two with locking brake.
AT NO TIME SHOULD THE INTERIOR OR
EXTERIOR BE STEAM CLEANED, HOSED
DOWN, OR FLOODED WITH WATER OR
LIQUID SOLUTION OF ANY KIND. DO NOT
USE WATER JET TO CLEAN.
S E V E R E D A M A G E O R E L E C T R I C A L H A Z A R D
C O U L D R E S U LT.
WARRANTY BECOMES VOID IF APPLIANCE IS FLOODED.
THIS UNIT HAS NOT BEEN
APPROVED FOR WARMING OF
BLOOD OR BLOOD PRODUCTS.