AERONEX® H SERIES GAS PURIFICATION SYSTEM

Safety Features

Feature

Description

PGPS4 EGPS4 EGPS8 EGPS12

Circuit breaker 

Provides additional electrical protection to the system and
includes a lock-out/tag-out.

N/A

Yes

Yes

Yes

Over temperature 
rise condition 

Monitored via thermocouple. Heaters sized to prevent runaway
conditions. As a secondary precautionary device, a high-
temperature hardware interlock is included on the EGPS4. 

Yes

Yes

Yes

Yes

Hydrogen 
detector

Triggers an EMO alarm and shuts down the system in the event 
of a gas leak of hydrogen inside system enclosure.

N/A

Yes

Yes

Yes

Rate-of-rise 
detector

If the detector senses a rapid increase in temperature inside the
system enclosure, an EMO alarm with be activated and shut
down the system.

N/A

Yes

Yes

Yes

EMO button 

When activated, power is removed from the cabinet. The system
shuts down. The front panel and controller remain powered. 

N/A

Yes

Yes

Yes

Remote EMO 

Provides input for remote EMO activation and an output for
remote signal of EMO condition. 

Yes

Yes

Yes

Yes

Remote alarm 

In the event of a minor alarm in the system not requiring an 
EMO shutdown, the system will send a signal to an external 
sensing device that alerts the facility of the alarm. 

Yes

Yes

Yes

Yes

Visual alarm 

In the event of an alarm, a detailed description of the alarm 
will be displayed in red on systems that include a touch screen. 
In the event of an alarm on a system with LEDs, a red LED 
indicator will activate. 

Yes

Yes

Yes

Yes

Audible alarm 

Alarm conditions result in an audible alarm. 

N/A

Yes

Yes

Yes

Static pressure
switch/door 
interlock

Monitors system for adequate static pressure. If enclosure 
ventilation falls below acceptable levels, an alarm is generated 
in the system; also serves as a door interlock to prevent the 
system from being used while the door is opened.

N/A

Yes

Yes

Yes

Isolated 
electrical 
enclosure 

Electronics are physically isolated from the main purifier cabinet in
an attached electrical enclosure in situations where high-voltage
lines are near potentially flammable gas. Low-voltage components
located inside the enclosure are SELV (safety extra low voltage).

N/A

Yes

Yes

Yes

Single or dual 
contained plumb-
ing interface

Allows end user to use either single or dual contained plumbing 
for installation.

Yes

Yes

Yes

Yes

Summary of Contents for AERONEX EGPS12H

Page 1: ...AERONEX H SERIES GAS PURIFICATION SYSTEM Continuous ultrapure H2 gas at a low cost of ownership ...

Page 2: ...to a process tool Applications Photolithography Metallorganic Chemical Vapor Deposition MOVCD Atomic Layer Deposition ALD Low Temperature Epitaxy LTE Other applications that require ultrapure hydrogen gas Features and Benefits Power failure will not damage the purification system Complete automatic operation saves time increases reliability Purifies to sub ppb part per billion and ppt part per tri...

Page 3: ...entation Diagram Models PGPS4H EGPS4H EGPS8H Process in R1 V2A V2B Purifier A Purifier B Moisture Sensor Process in Process out Purifier A Purifier B Regen vent Rotameter Optional Moisture Sensor Regen input V8 optional V1B V2B V1A R1 PT2 PT1 V2A V7A V7B V5 V6 V3B V4B V3A V4A Model EGPS12H ...

Page 4: ...rm In the event of a minor alarm in the system not requiring an EMO shutdown the system will send a signal to an external sensing device that alerts the facility of the alarm Yes Yes Yes Yes Visual alarm In the event of an alarm a detailed description of the alarm will be displayed in red on systems that include a touch screen In the event of an alarm on a system with LEDs a red LED indicator will...

Page 5: ...rocess gas output Mechanical connection 1 4 face seal 1 4 tube stub 1 2 tube stub 3 4 tube stub Ventilation Mechanical connection N A 4 duct Exhaust flow N A 50 CFM 100 CFM 65 CFM Power requirements Mechanical connection Standard terminal Power requirements 200 240 VAC Power consumption 50W at idle and process mode 300W during regen 50W at idle and process mode 800W during regen 50W at idle and pr...

Page 6: ...vent Exhausts regen gas 6 Process gas input Inlet gas not purified 7 Process gas output Outlet gas purified Panel Information This equipment is not enclosed It is the user s responsibility to ensure that it is installed in compliance with local safety requirements for gas equipment The PGPS4 is designed using SEMI S2 guidelines for gas equip ment enclosures Because it is a subsystem it must be cer...

Page 7: ...ace Allows for remote alarm input and output with female 15 pin DB connector 7 Regen gas vent Exhausts regen gas 8 Exhaust vent Allows ventilation 9 Process gas input Inlet gas not purified 10 Process gas output Outlet gas purified 11 Reference Atmospheric reference for the internal enclosure flow sensor 12 Instrument air Supplies gas to the air operated control valves 13 A C power input Power con...

Page 8: ...When activated power is removed from the cabinet The system shuts down Front panel and controller remain powered 5 Remote alarm interface Allows for remote alarm input and output with female 15 pin DB connector 6 Instrument air Supplies gas to the air operated control valves 7 Regen gas vent Exhaust regen gas 8 Exhaust vent Allows ventilation 9 Process gas input Inlet gas not purified 10 Circuit b...

Page 9: ... Fire sensor located below cover 10 Exhaust vent Allows ventilation 11 Process gas output Outlet gas purified 12 Gas monitor located below cover 13 Regen gas vent Exhausts regen gas 14 Instrument air Supplies gas to the air operated control valves 15 Reference Atmospheric reference for the internal enclosure flow sensor 16 Circuit breaker Provides additional electrical protection to the system and...

Page 10: ...up to 300 SLM up to 1000 SLM PGPS4H EGPS4H EGPS8H EGPS12H PGPS4HA EGPS4HA EGPS8HA EGPS12HA PGPS4HAT EGPS4HAT EGPS8HAT EGPS12HAT PGPS4HM EGPS4HT EGPS8HM EGPS12HM PGPS4HMT EGPS8HMT EGPS12HMT PGPS4HT EGPS8HT EGPS12HT Automatic Manual Part Bypass Bypass Moisture Number Manifold Manifold Indicator PGPS4H PGPS4HA Yes PGPS4HAT Yes Yes PGPS4HM Yes PGPS4HMT Yes Yes PGPS4HT Yes EGPS4H EGPS4HA Yes EGPS4HAT Y...

Page 11: ...onal U S Patent 6 361 696 2004 2006 Entegris Inc All rights reserved Printed in USA 4507 2676ENT 1206 For More Information Please call your Regional Customer Service Center today to learn what Entegris can do for you Visit www entegris com and select Regional Customer Service Centers for the center nearest you Terms and Conditions of Sale All purchases are subject to Entegris Terms and Conditions ...

Reviews: