96
1600751-4D
13.13 Vacuum pulsed mode
p
Vacuum pressure
r
Range
t
h
Hold time
13.14 Ultrasound
Ultrasound parameters:
f
p
Pulse frequency, 100 Hz fixed
t
Pulse duration, set by duty cycle
5
–
80 % and 100 % (100% is
continuous wave)
t
0.5
–
8 ms
RTPA 20
–
1.25
14 Specifications
14.1 Ultrasound parameters
See fig 13.14 for details.
Generator
Peak output Amplitude:
Duty cycle 5
–
50 %
: 0
–
3 W/cm²
Duty cycle 80 %
: 0
–
2.5 W/cm²
Duty cycle 100 %
: 0
–
2 W/cm² (continuous wave)
Peak output power for 5 cm² applicator:
Duty cycle 5
–
50 %
: 0
–
15 W
Duty cycle 80 %
: 0
–
12 W
Duty cycle 100 %
: 0
–
10 W (continuous wave)
Peak output power for 0.8 cm² applicator:
Duty cycle 5
–
50 %
: 0
–
2.4 W
Duty cycle 80 %
: 0
–
2 W
Duty cycle 100 %
: 0
–
1.6 W (continuous wave)
Output meter uncertainty
: ± 20 % for any output above 10 % of maximum
Pulse frequency
: 100 Hz ± 1 % fixed
Duty cycle
: 5
–
80 % and 100 % (100 % is continuous wave)
Pulse duration
: 0.5
–
8 ms ± 10 % (set by duty cycle)
Temporal Peak to Average Ratio (RTPA):20
–
1.25 ± 10 % (set by duty cycle)
Treatment timer
: 0 - 30 min ± 0.1 min, linked to contact control
Contact control level
: 65%
5 cm² Applicator
Ultrasound frequency:
1 MHz
: 0.98 MHz ± 2 %
3 MHz
: 3.1 MHz ± 2 %
Summary of Contents for 1600905
Page 1: ...6 series User Manual ...
Page 3: ...6 series User Manual ...
Page 4: ......
Page 5: ...3 1600751 4D ...
Page 90: ...88 1600751 4D ...
Page 92: ...90 1600751 4D 13 6 3 LP 13 6 4 CP 13 6 5 CPid ...
Page 95: ...93 1600751 4D 13 7 7 IG150 Monophasic Envelope 13 7 8 IG150 Biphasic Envelope ...