Page 69 of 72
EN109-1498751-47 IFU
Environmental conditions
Environmental conditions for transport and storage
Environmental temperature:
–
20° to +70° C
Relative humidity:
10 to 90% (non-condensing)
Atmospheric pressure:
500 to 1060 hPa
Environmental conditions normal use
Environmental temperature:
10° to 40° C
Relative humidity:
10 to 90 % (non-condensing)
Atmospheric pressure:
500 to 1060 hPa
EMC details
Medical electrical devices such as the 4-series are subject to special precautions with regard to
electromagnetic compatibility (EMC) and must be installed and commissioned in accordance
with the EMC advice given in the instructions for use.
Portable and mobile RF communication systems (e.g. mobile phones) may interfere with
medical electrical devices.
The 4-series should only be operated with the original power cable specified in the list of
contents delivered. Operating the device with any other power cable can lead to increased
emissions or reduced interference immunity of the device.
Use of this equipment adjacent to or stacked with other equipment should be avoided because
it could result in improper operation. If such use is necessary, this equipment and the other
equipment should be observed to verify that they are operating normally.
Use of accessories, transducers and cables other than those specified or provided by the
manufacturer of this equipment could result in increased electromagnetic emissions or
decreased electromagnetic immunity of this equipment and result in improper operation.
The main features of the 4-series devices are as follows: interference-free delivery of shockwaves,
interference-free control of all functions. Uninterrupted operation is not required with the use
intended.