Page 64 of 72
EN109-1498751-47 IFU
Cleaning the
water reservoir
and hoses
•
Detach the vacuum cups from the vacuum cables.
•
Place a container filled with a cleaning liquid (*)
below the system.
•
Place the peripheral ends of the cables in the container.
•
Go to System Settings and select Tank Cleaning.
•
The water reservoir will be filled with the cleaning liquid until the water
reservoir is full.
•
Empty the water reservoir as described below (
water reservoir full
).
(*) The following registered products may be used for disinfecting the water reservoir:
BAKTOLAN to 5%, CHINOSOL to 1%, CHLORAMIN solution, ELMOCID Gamma to 2%, MEFAROL to 1%,
MERCKOJOD to 1%, MERFEN, PERHYDROL, PERODIN, SAGROTAN to 2%, ZEPHIROL to 5%.
Troubleshooting
When the device is turned on, it will first execute a self-test. When an error is detected, both
during the self-test and during normal operation, a pop-up screen will appear on the display.
When the error is displayed, all outputs will be disabled. When this situation occurs, remove all
cables, and switch the apparatus off and on again. When the error re-appears, stop using the
device and contact your supplier.
Errors and warnings indicate an internal problem with the device that must be tested by a field
service technician certified by Enraf-Nonius B.V. before any further operation or use of the
system. Use of a device that indicates an error or warning may pose a risk of injury to the
patient, user, or extensive internal damage to the system. In case of display failure or other
obvious defects, unplug the device immediately and notify a certified service technician.
Patient circuit interrupted
There is insufficient or no output current available at the outputs. Possible causes:
•
Poor electrical contact or broken patient cable.
•
Insufficiently moistened sponges. If necessary, use a saline solution to improve the electrical
conductivity of the water.
•
Too high current amplitude with self-adhesive electrodes. Try to continue the treatment with
flexible rubber electrodes.
If the problem occurs in CC mode, the current amplitude will ramp down to 0 and will have to be
readjusted when the problem has been solved.
If none of the above scenarios appear to be the problem, stop using the device and contact your
supplier.
Battery low
The battery is insufficiently charged to complete the treatment at the currently set therapy levels.
Reduce the therapy levels or connect the apparatus to the mains supply.
Water reservoir full
The water separation tank of the Vacotron is full. Continue the treatment with the standard electrodes
or empty the tank as follows:
•
Set power line switch [1] Off (0).
•
Detach the hose from the upper hose nipple [15] and empty the reservoir.
•
Reattach the hose to the hose nipple.
•
Set power line switch [1] On (1).