Enraf-Nonius B.V.
Curapuls 670 Service Manual
1403773
Maintenance
June 1997
4-13
-
Remove the connection between TP16 and TP17 and connect an ampère-meter
(DC-mode) instead of it, see appendix B, fig.8.
-
Place the test jumper at ST4, see appendix B, fig.8.
See appendix B, fig. 8 for lay-out of the adjustment / connection points.
Warning:
At the mounting base of the power transistors near jumper ST4 voltages
up to 400V. are present. Do not touch when the Curapuls 670 is
switched on!
-
Switch on the Curapuls 670 and select test 38, par. 4.6.4. (see par. 4.3.2 for entering
the service mode).
-
Choose subselection 0 (300W HF output) and press the start / stop key,
C162
Adjust C162 until the voltmeter gives a maximum reading.
L20
Adjust with L20 until the voltmeter indicates 2.45 volt ± 0.05 volt.
0.7A
Check that the current is less than 0.7 Ampere,
L15
If the current is more than 0.7A, adjust by pressing L15.
(if this does not work; an alternative can be to compensate with L23). Check if
the voltmeter is still at maximum reading.
-
press the start / stop key
ST4
Remove the jumper from ST4.
-
Choose subselection 1 (200W HF output) and press the start / stop key,
P1
Adjust P1 until the voltmeter reads 2.00 volt ± 0.1 volt.
0.6A
Check that the current is less than 0.6 A.
The calibration of the Curapuls 670 is completed.
4.6.5
Frequency / SWR adjustment Circuplode (Test 39)
-
Connect the Circuplode to output channel A
-
Select test 39, par. 4.6.3 (see par. 4.3.2 for entering the service mode).
-
Adjust the Circuplode until the Result dec. display reads 27.14
For technical information about adjustment of the Circuplode(s) see:
Service manual “Curamed and Curapuls accessories” (Part number 1419.781).
Summary of Contents for 1403773
Page 1: ...Curapuls 670 SERVICE MANUAL Part number 1403773 ...
Page 2: ......
Page 6: ...Enraf Nonius B V Curapuls 670 Service Manual 1403773 Index June 1997 4 ...
Page 16: ...Enraf Nonius B V Curapuls 670 Service Manual 1403773 Test equipment and tools June 1997 3 2 ...
Page 30: ...Enraf Nonius B V Curapuls 670 Service Manual 1403773 Maintenance June 1997 4 14 ...
Page 32: ...Enraf Nonius B V Curapuls 670 Service Manual 1403773 Storage and transport June 1997 5 2 ...
Page 48: ...Enraf Nonius B V Curapuls 670 Service Manual 1403773 Appendix B June 1997 B 12 ...
Page 56: ......
Page 58: ...No SI 670 02 Oct 1998 Page 2 of 2 Figure 1 Detail of the main PC Board ...