HEAD AND FACE

For common head and cranial ailments associated with conditions like 
headaches, migraines, TMJ disorders, stress, depression, anxiety, and 
other common head and cranial ailments:

1.  We recommend using the e-Pulse® Auricular Ear Clips (sold    
 

separately) and SELECT a CES therapy such as MICROCURRENT   

 

(MODE 11 and 12) and set the DURATION to at least 20 to  

 

 

30 MINUTES. 

2.  You can also use small electrode pads placed over your  

 

 

Tempomandibular joint (TMJ) to treat pain associated with TMJ    

 

disorders, and we suggest using the same MICROCURRENT therapy  

 

for at least 20 to 30 MINUTES.

 CLEANING AND STORAGE

The unit is designed for repeated use over time. The pads will last up to 
180 uses, or 5 months (based on daily use). Here are important cleaning 
and storage instructions:

CLEANING AND STORING THE PADS:

1.  Turn your unit OFF.
2.  DISCONNECT the electrode wires from the unit.
3.  DISCONNECT the massage pads from the electrode wires.
4.  Gently rub a few drops of water on the adhesive gel using your   
 

fingertips until all impurities have been rubbed away.

5.  You can also wash the pads under slow running cold water  

 

 

for serveral seconds while gently rubbing the adhesive gel with   

 

your fingertips.

6.  DO NOT use a sponge, cloth, detergents, chemicals, or soap.
7.  We recommend washing the pads every 10 to 15 uses.  

 

 

Do not over-wash your pads.

8.  Place the massage pads on the pad holder by sticking them to the  
 

sides of the holder and let dry completely.

9.  DO NOT wipe with a tissue or cloth.
10. Do note keep in areas subject to direct sunlight, high or low  

 

 

temperatures, high humidity, near fire, or electricity.

11. 

Replacement pads can be purchased online at  

 

 

 

www.epulsemassage.com or by calling (844)-4-EPULSE (37-8573).

CLEANING AND STORING THE UNIT:

1.  Turn the unit OFF.
2.  DISCONNECT the electrode wires from the unit and leave the    
 

electrode wires connected to the pads for convenience.

3.  STORE the pads and wires on the pad holder by sticking the pads to  
 

the flat side of the holder; then wrap the wires around the groove on  

 

the edge of the pad holder.

4.  Gently wipe the unit with a microfiber cloth.
5.  DO NOT use water or chemicals.
6.  STORE the unit in a safe and convenient place away from the reach 
 

of children.

7.  Do note keep in areas subject to direct sunlight, high or low  

 

 

temperatures, high humidity, near fire, or electricity.

8.  DO NOT wrap the wires around the unit because it may damage  
 

the wires.

TROUBLESHOOTING GUIDE

PROBLEM 

POSSIBLE CAUSE 

SOLUTION 

I can’t feel anything 

 

Ba

ery is dead 

 

Massage pad is dry, 

dirty, or does not s

ck 

 

Damaged massage pad 

and/or electrode wire 

 

Using only 1 pad 

 

Bad Channel port 

 

Skin is too dirty, oily, or 

hairy 

 

Recharge the ba

ery 

 

Clean or replace the 

massage pad(s) 

 

Replace electrode 

wire(s) and/or 

massage pad(s) 

 

ALWAYS USE 2 PADS 

 

Test Channel port A 

with pads on. Then 

test Channel B with 

pads on. If one of the 

Channels works but 

one doesn’t, you may 

have a bad port. 

Contact Customer 

Support. 

 

Wash skin thoroughly 

to remove dirt, oils, 

lo

ons, and any 

impuri

es. Shave if 

needed. 

Irregular pulse 

 

Poor connec

on 

 

Damaged massage 

pad and/or electrode 

wire 

 

Massage pad is dry, 

dirty, or does not 

s

ck 

 

Skin is too dirty, oily, 

or hairy 

 

Pads overlapping 

 

Reconnect unit to 

electrode wire 

 

Replace electrode 

wire and/or massage 

pad(s) 

 

Wash skin thoroughly 

to remove dirt, oils, 

lo

ons, and any 

impuri

es. Shave if 

needed. 

 

Make sure pads do 

not overlap.  

Summary of Contents for e-Pulse Ultra

Page 1: ... 07 20 to 30 Minutes WORKOUT Mode 13 14 10 Minutes RUSSIAN STIM Mode 16 10 Minutes MICROCURRENT Mode 11 12 20 to 30 Minutes KNEADING Mode 05 10 Minutes CHOPPING Mode 01 20 to 30 Minutes WORKOUT Mode 13 14 10 Minutes RUSSIAN STIM Mode 16 10 Minutes ARM PAIN ACUPUNCTURE Mode 03 10 Minutes SWEDISH Mode 10 20 to 30 Minutes WORKOUT Mode 13 and 14 10 Minutes RUSSIAN STIM Mode 16 10 Minutes HEAD PAIN MIC...

Page 2: ...EMOVE PLASTIC FILM from PADS 3 PLACE PADS ON SKIN Refer to the Pad Placement guide 1 SNAP ELECTRODE WIRE to MASSAGE PADS 11 PRESS and HOLD the CENTER BUTTON for 3 SECONDS to LOCK UNLOCK the device 10 PRESS the D BUTTON to ADJUST DURATION of the therapy 9 PRESS the CENTER BUTTON to toggle between CHANNEL A CHANNEL B and DUAL CHANNEL CONTROL REPEAT STEPS 6 9 to select different THERAPIES for CHANNEL...

Page 3: ...ww epulsemassage com for more information on e Pulse products and offers Manufactured for and distributed by Enovative Technologies LLC 2016 Enovative Technologies LLC Stronger pulse Maximum relief USER MANUAL ...

Page 4: ...o help with muscle joint or bone problems that occur with conditions such as sciatica lower back pain neck pain shoulder pain knee discomfort tendinitis bursitis or with illnesses such as osteoarthritis or fibromyalgia WHAT IS CES Cranial Electrotherapy Stimulation CES is a form of non invasive brain stimulation that applies a small pulsed electric current across a person s head to help relieve st...

Page 5: ...s product on your chest Electric pulses may cause disturbances in your heart rhythm and could be harmful 7 Do not use this product on your face or around the mouth 8 Do not use in or around your groin area 9 Do not place pads directly over the spine 10 Do not use over open wounds sores rashes or over swollen irritated infected skin or over cancerous lesions 11 Do not use while driving a vehicle op...

Page 6: ...aged or worn out massage pads may cause discomfort irritation or minor skin burns 8 We recommend that you only use accessories sold by Enovative Technologies 9 Using this device near short wave or microwave therapy equipment may cause an unstable power output 10 Keep your e Pulse Ultra clean by wiping with a microfiber cloth Do not use water 11 Do not disassemble the unit The warranty will be void...

Page 7: ...t screen Six massage modes combined into 16 unique therapies 2 Swedish 2 Acupuncture 2 Shiatzu 2 Thai 2 Microcurrent 2 Auricular 2 Reflexology and 2 Workout 20 levels of intensity for soft and strong massages Independently controlled dual outputs support up to four pads at once Adjustable timer 10 60 minutes Extremely portable lightweight unit with small dimensions Long lasting rechargeable lithiu...

Page 8: ...PRODUCT DIAGRAM 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 T D ...

Page 9: ...rapy Icon Indicator 16 Active therapy indicator 17 First therapy variation indicator 18 Second therapy variation indicator 19 Active Channel A and Channel B indicator 20 Channel selection indicator 21 Intensity level indicator 22 D button for adjusting Duration of therapy 23 button for increasing therapy intensity 24 Therapy T button for selecting your desired therapy 25 Channel B input THERAPIES ...

Page 10: ...f your unit STEP 2 SNAP ELECTRODE WIRE TO MASSAGE PADS Connect the two ends of the electrode wire to the massage pads by snapping the round plugs on the pins attached to the cloth side of the massage pads NOTE Please make sure to clean your skin thoroughly to remove any oils dirt or impurities as these may cause the adhesive gel to fail during therapy Pads may not stick if the skin has too much ha...

Page 11: ...or across the targeted area such as your ankle across the length of your shin around your calf knee hamstring or across your thigh 2 Treat multiple areas simultaneously using four pads or use 6 pads with the optional dual connector 3 Place pads vertically when pain extends across a significant distance of your legs or in either side when treating a smaller area NECK Warning We recommend using smal...

Page 12: ...separately 1 Carefully place the massage pads on either side or across the length of your forearm around your elbow or across your bicep and or triceps 2 Treat multiple areas simultaneously using four pads or use 6 pads with the optional dual connector 3 Place pads vertically when pain extends across a significant distance of your arms or on either side when treating a smaller area HEAD Warning Fo...

Page 13: ... 5 PRESS the CENTER button to SELECT CHANNEL A ONLY the A and B CHANNEL indicator inside the middle circle tells you which CHANNEL you are selecting 6 At this point you can either ADJUST the INTENSITY of your therapy or ASSIGN a THERAPY to CHANNEL A ONLY 7 PRESS or to modify the intensity of your therapy 8 PRESS the T button to ASSIGN a therapy to CHANNEL A ONLY continue pressing the T button to c...

Page 14: ... you are looking to build or rehabilitate a group of muscles use EMS therapy WORKOUT MODE 13 and 14 or RUSSIAN STIM MODE 16 Set the DURATION to at least 10 MINUTES and increase the INTENSITY to as high as you can tolerate BACK For common back problems associated with conditions such as fibromyalgia low back pain sciatica nerve bone and muscular injuries herniated or degenerative disc disease arthr...

Page 15: ...er great therapy for shoulder pain is CHOPPING MODE 01 and set the DURATION to 20 or 30 MINUTES 3 If you are looking to build muscle we recommend EMS therapy such as WORKOUT MODE 13 or 14 or RUSSIAN STIM mode 16 Set the DURATION to at least 10 MINUTES and increase the INTENSITY to as high as you can tolerate ARMS For common arm problems associated with conditions such as fibromyalgia nerve bone an...

Page 16: ...and storage instructions CLEANING AND STORING THE PADS 1 Turn your unit OFF 2 DISCONNECT the electrode wires from the unit 3 DISCONNECT the massage pads from the electrode wires 4 Gently rub a few drops of water on the adhesive gel using your fingertips until all impurities have been rubbed away 5 You can also wash the pads under slow running cold water for serveral seconds while gently rubbing th...

Page 17: ... CAUSE SOLUTION I can t feel anything Ba ery is dead Massage pad is dry dirty or does not s ck Damaged massage pad and or electrode wire Using only 1 pad Bad Channel port Skin is too dirty oily or hairy Recharge the ba ery Clean or replace the massage pad s Replace electrode wire s and or massage pad s ALWAYS USE 2 PADS Test Channel port A with pads on Then test Channel B with pads on If one of th...

Page 18: ...s about our warranty policy Pulse is not strong enough Low ba ery Damaged massage pad and or electrode wire Massage pad is dry dirty or does not s ck Using only 1 pad Recharge ba ery Replace electrode wire and or massage pad Recharge the ba ery Clean or replace the massage pad ALWAYS USE 2 PADS Pulse surges unexpectedly Poor connec on Massage pad is dry dirty or does not s ck Damaged massage pad s...