Installation
Installation and Operation Manual – AQUATEC 1000
14
Doc.-ID: AQUA_16012020_EN
2.9
Adjusting the V-shield / filter head
Before installing the probe,
the direction of flue gas flow must be determined
and the filter head assembly turned
to such a position that the V-shield faces the oncoming flue gas. The filter head can be turned freely a full 360° for
this purpose, by loosening the counter nut, loosening the Allen screw and rotating the filter head / V-shield to the
required position, and then tightening the Allen screw and counter nut.
Flue gas direction
V-shield (KES-132X/KES-200X)
V-shield ( KES500X)
Screws to secure filter head
Filter head
Rotation of filter head
Probe
Figure 11 -
Adjustment of the V-shield
Loosen the locknut
Loosen the locknut
Adjust the filter head.
Adjust the position of the
shield against the flow direction
Tighten the locknut
Tighten the locknut
Figure 12 -
Adjustment the fitter head
1
2
3
4