4.2 Numeric and Character Indications

4.3 Protection Levels

Each parameter of the EDI62S has an individual protection level, and by setting the protection level of the condition data, you can set an access level.
(For the protect level of each parameter, see the P.L. column of the tables in Section 4.4.)
The higher the protection level is, the less the number of settable parameters will be. If you set the protection level to the strictest LV3, you can change
the protection level only, and all the other parameters may not be changed. (No comparator judgment value can be changed using the jog switches in
this case.)
*The protection level set at the time of shipment is LV1. (Settings of display colors, scaling and judgment-related values only are
available.)

5.6 Calibration Data Setting Examples

How to calibrate the zero input (ZRIN) value and the span input (SPIN) value using an actual load or the equivalent input

Push the center of Jog 1 to choose the digit you want 
to change, and move it up or down to change the 
value. In the left figure, pushing the shift while the 
least significant digit of 

1.000

 is selected, causes 

the lower digits to also be displayed.

Push and hold the Shift key, then the current 
measured value will be read as the setting.

Equivalent input calibration

Actual load calibration

Push and hold

Mode

&

* The EDI62S series does not add the zero input value to the span input value.
    Namely, the gradient and offset values for y=ax+b are obtained as follows:

Gradient (a) = 

(SPAN - ZERO)

(SPIN - ZRIN)

Offset (b) = 

(SPAN - ZERO) ×ZRIN

(SPIN - ZRIN)

- ZERO

1

Calibration data pattern selection
Select a pattern to set.
* Patterns cannot be selected when the number of available patterns is set to 1 using condition data.

1

1

1

Sensor power supply selection

1

1

1

1

1

Zero input value setting

1

Mode

1 Select a digit to set.

1

1

Change the value for the selected digit 
(equivalent input calibration).

the current measurement as the setting (actual load calibration).

Zero indication value setting

1

Mode

1 Select a digit to set.

1

1 Change the value for the selected digit.

1

1

1

Mode

1

1

1

Change the value for the selected digit 
(equivalent input calibration).

1

1

Mode

1 Select a digit to set.

1

1 Change the value for the selected digit.

Select a digit to set.

Operation Method

Press the Shift key for prolonged time to automatically read

the current measurement as the setting (actual load calibration).

Press the Shift key for prolonged time to automatically read

How to set a low layer value (Condition data/scaling data/comparator data)

Mode

Mode

Enter

Select the name of a parameter you want to change by pushing the Increment key.

Use the Shift key to choose the digit you want to change, and push the Increment key to set the value.

If the parameter requires a value entry:

Use the Increment key to choose a selection.

If the parameter requires a selection:

Use the Mode key to shift from the value change mode to the menu selection mode.

Use the Increment key to select the menu name of the parameter you want to change or return to 
measurement by pushing the Enter key.

To change the selected parameter, use the Mode key to move from the menu selection mode to the value 
change mode.

&

or