ENERGY SUPPORT CORPORATION DTF-201 Operation Manual Download Page 1

 

 

Manual No.3156E R4 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Operation Manual for 

DTF-201 

Oxygen Analyzer 

(% Range) 

 

 

RX-62230

-A

*****

 

 

Summary of Contents for DTF-201

Page 1: ...Manual No 3156E R4 Operation Manual for DTF 201 Oxygen Analyzer Range RX 62230 A...

Page 2: ......

Page 3: ...ar heat resistant gloves or the like and be careful to avoid burns 3 An electric shock warning mark shown on the right is attached near the power supply where there is a danger of electric shock If th...

Page 4: ...ating state of the analyzer However the following cases shall be excluded from the scope of guarantee even if they occur in the above mentioned term of guarantee 1 Failure generated due to operation e...

Page 5: ...1 Installation Conditions 5 3 2 Installation Method 6 3 3 Piping and Wiring Methods 6 4 Operation 8 4 1 Operation Preparation 8 4 2 Start up 9 4 3 Stop Procedure 10 4 4 Procedure During Operataio 10 4...

Page 6: ...nalyzer Do not use sample gas containing corrosive gases F HF CL2 HCL SO2 H2S etc or poisonous materials Si Pb P Zn Sn As etc These can shorten the life of the sensor Do not use sample gas containing...

Page 7: ...ell When current is applied between the electrodes the oxygen inside the gas detection chamber is transferred to the reference oxygen chamber so that the oxygen concentration in the reference oxygen c...

Page 8: ...O2 concentration Gas Gas diffusion holes Gas Heater Pumping cell Gas detection chamber Sensing cell Reference oxygen chamber Heater temperature control Pumping current Output signal Pumping current co...

Page 9: ...n data and error Display 2 3 digits for display of range concentration alarm and channel range lamp Lit in measurement mode ppm range lamp Lit in ppm measurement mode Key for calibration and data sett...

Page 10: ...ht The ambient temperature is between 10 and 50 with little variation in temperature There is little humidity and dust The location is not exposed rainfall There is little mechanical vibration There a...

Page 11: ...ng Arrangement Piping between the probe generator sensor and the air selector unit for supplying calibration gas must be connected Pipes consist of the following Calibration gas pipe Ejector air pipe...

Page 12: ...of oxygen concentration corresponding to selected range Non insulated output load resistance 600 or less OUT2 0 to 5 VDC output of oxygen concentration corresponding to selected range Non insulated o...

Page 13: ...n Start pre operation inspection Connect it pre operation inspection completed Is the power cord connected properly Is the connecting cable connected properly NO Is the analyzer input output wiring co...

Page 14: ...1 When range No 2 is selected 2 When range No 3 is selected 3 When range No 4 is selected 4 Checking and changing of each range is performed using the receiver setting data CH030 037 If the concentra...

Page 15: ...xygen analyzer output signals for control purposes always apply the control release device before operating the keys Key Functions key Press this key to shift from the O2 Concentration display to the...

Page 16: ...CH No Press or to change the data displayed on Display 1 Press to save the data in the receiver memory Press to return to the O2 Concentration display Key operation is finished Using steps above you...

Page 17: ...PPORT CORP 016 Primary delay time sec 0 99 0 017 Data for setting checking by ENERGY SUPPORT CORP 018 WET DRY operation 0 WET 1 DRY gas fuel 2 DRY solid and liquid fuel See Specifications Manual 019 0...

Page 18: ...to 142 Data for setting checking by ENERGY SUPPORT CORP 143 Linearization table Specific values for each sensor See inspection data 144 179 Data for setting checking by ENERGY SUPPORT CORP 180 Calibra...

Page 19: ...range switching 210 219 Data for setting checking by ENERGY SUPPORT CORP 220 O2 concentration max alarm setting value 0 0 9999 0 If 5 and 6 are set for CH200 203 set your desired limit 221 O2 concentr...

Page 20: ...calibration 0 Manual 1 Semi Auto 2 Auto 276 Calibration mode 0 Air 1 point calibration 1 Air Zero 2 point calibration 2 Air Span 2 point calibration 3 Air Zero Span 3 point calibration See Specificati...

Page 21: ...e span value setting is CH031 unit setting is CH035 For output range No 3 range span value setting is CH032 unit setting is CH036 For output range No 4 range span value setting is CH033 unit setting i...

Page 22: ...ransmitter sensor unit at 1 3L min and it stabilizes indication Air point is calibrated in a method to show in c e of c Input to CH121 for span gas concentration The concentration of span gas is to us...

Page 23: ...s not changed steps a c and e are not required Input the concentration after the change by the operation of c when you change the concentration of span gas 4 5 Operation for when an Error Occurs If an...

Page 24: ...tput becomes DRY output by the above 3 Output hold setting If an error occurs for the sensor or receiver while CH023 is selected this sets how the concentration output signal is to be held The relatio...

Page 25: ...s called up the display shows cAL 2 When you press the key 0 is displayed on display 1 Each time you press the key display 1 can be switched to 0 and 100 When 0 is displayed output ZERO adjustment is...

Page 26: ...s Contact output operation CH 204 207 input data 0 NO setting 1 NC setting 0 No contact output OFF ON 1 Analyzer error During warming up measurement OFF ON Upon analyzer error ON OFF 2 Range echo 2 ra...

Page 27: ...measurement Make sure you thoroughly understand the procedure before performing maintenance Sensor replacement and pump maintenance cautions To prevent gas intoxication or oxygen de ciency before you...

Page 28: ...eating heat is carried out Gas calibration error Perform gas calibration Sensor deterioration Replace the sensor Analyzer output display value is zero Flammable gas included in sample gas Eliminate fl...

Page 29: ...12 displayed E 13 displayed Sensor failure Replace sensor Receiver problem Request repair by ENERGY SUPPORT CORP E 14 displayed Heater terminal S S release Check for wiring E 21 displayed Data is inp...

Page 30: ...less Linearity 1 FS 0 5 O2 1 FS or less Response Time 10 sec or less 90 response for calibration gas switching Warming up time Approx 3 min Ambient temperature 0 50 Humidity 90 RH or less Power AC85...

Page 31: ...rm display and contact output Error is displayed Error number Indication does not change except for zero span point Error lamp is lit Indication overshoots the scale span side Error contact output is...

Page 32: ...livery date instrument Operation conditions Measuring range Gas temperature Measuring concentration Gas pressure Situations of troubles Date and time of event occurrence At heating up steady running c...

Page 33: ......

Page 34: ...Description in this operation manual is subject to changes without prior notice For advice or detail of our products and services contact the Technical Service Team Industrial System Division...