background image

Commissioning

Turbimax CUS71D

20

Hauser

6.2  

Post-connection check

Device condition and specifications

Notes

Are the sensor, assembly, or cables free from damage on the outside?

Visual inspection

Electrical connection

Notes

Are the mounted cables strain-relieved and not twisted?

Is a sufficient length of the cable cores stripped, and are the cores
positioned in the terminal correctly?

Check the fit (by pulling gently)

Are all the screw terminals properly tightened?

Tighten

Are all cable entries mounted, tightened and leak-tight?

For lateral cable entries, make sure the cables
loop downwards to allow water to drip off

Are all cable entries installed downwards or mounted laterally?

7  

Commissioning

7.1  

Preparatory steps

7.1.1  

Factory settings

The sensor is preconfigured on leaving the factory. Once the installation parameters have

been adjusted, the sensor is ready for use in a wide range of applications without further

adjustments. The factory settings cannot be deleted and can be restored at any time.

Enter the installation parameters at the Liquiline CM44x transmitter.

For more information, please refer to the Operating Instructions of the transmitter.

8  

Operation

8.1  

Sensor monitoring

The acoustic signals are continuously monitored and analyzed for plausibility. If

inconsistencies occur, an error message is output via the transmitter.
In addition, the following fault conditions are detected by the sensor diagnostics of Liquiline

CM44x:

• Implausibly high or low measured values

• Disturbed regulation due to incorrect measured values

Summary of Contents for Turbimax CUS71D

Page 1: ...Products Solutions Services Operating Instructions Turbimax CUS71D Ultrasonic interface sensor BA00490C 07 EN 06 18 71415440 2018 05 01...

Page 2: ......

Page 3: ...stallation conditions 11 5 2 Mounting the sensor 14 5 3 Post installation check 17 6 Electrical connection 17 6 1 Connecting the sensor 17 6 2 Post connection check 20 7 Commissioning 20 7 1 Preparato...

Page 4: ...situation Failure to avoid the dangerous situation can result in a fatal or serious injury LCAUTION Causes consequences If necessary Consequences of non compliance if applicable Corrective action This...

Page 5: ...must be carried out only directly at the manufacturer s site or by the service organization 2 2 Designated use CUS71D is a sensor designed for interface measurement in water and wastewater The sensor...

Page 6: ...allation guidelines Local standards and regulations Electromagnetic compatibility The product has been tested for electromagnetic compatibility in accordance with the applicable European standards for...

Page 7: ...peration 2 5 Product safety The product is designed to meet state of the art safety requirements has been tested and left the factory in a condition in which it is safe to operate The relevant regulat...

Page 8: ...ultrasonic signal needs to reach the solid particles in the separation zone and return to the receiver The separation zone is computed after the maximum slope and the maximum signal amplitude A sensor...

Page 9: ...s On the nameplate In the delivery papers Obtaining information on the product 1 Go to www endress com 2 Call up the site search magnifying glass 3 Enter a valid serial number 4 Search The product str...

Page 10: ...irectives The manufacturer confirms successful testing of the product by affixing to it the mark 4 4 2 EAC The product has been certified according to guidelines TP TC 004 2011 and TP TC 020 2011 whic...

Page 11: ...D Sensor offset E Basin depth F Opening angle of ultrasonic cone 6 Installation instructions Find a suitable installation position for the sensor in the basin Take the following points into considerat...

Page 12: ...r or foam formation occur e g inlet 3 Using a dip pipe install the sensor at least 20 cm 0 66 ft beneath the surface of the water 5 1 2 Dimensions 74 2 9 75 3 0 33 7 1 3 20 5 0 8 17 0 7 15 0 6 G1 NPT...

Page 13: ...1 2 3 4 5 11 12 13 14 A B A0031579 4 Basin configuration in circular clarifiers A View from above B Cross section 1 Surface skimmer 11 Sensor 2 Bridge walkway 12 Rail 3 Sensor position range 13 Surfac...

Page 14: ...g system A complete measuring system comprises Turbimax CUS71D ultrasonic sensor Liquiline CM44x multi channel transmitter and is optionally delivered with the following accessories A weather protecti...

Page 15: ...5 Ultrasonic sensor with basin holder system and multi channel transmitter 1 Flexdip CYH112 holder 2 Liquiline CM44x multi channel transmitter 3 Protective cover 4 Flexdip CYA112 assembly 5 Turbimax...

Page 16: ...dip CYH112 holder cross clamp 2 Flexdip CYH112 holder pendulum holder 3 Flexdip CYA112 assembly with CUS71D 4 PVC sensor protector 5 Splash protection cap The PVC sensor protector protects the ultraso...

Page 17: ...Electrical connection LWARNING Device is live Incorrect connection may result in injury or death The electrical connection may be performed only by an electrical technician The electrical technician...

Page 18: ...8 87 32 31 32 31 PK GY GN YE Sensor PE A0034802 7 Sensor connection The maximum cable length is 100 m 328 ft 3 Use the following accessories to extend the sensor cable if necessary CYK11 measuring cab...

Page 19: ...s not necessarily correspond to the original cable supplied 1 2 3 8 Terminated cable 1 Outer shield exposed 2 Cable cores with ferrules 3 Cable sheath insulation 4 9 Insert the cable 4 Grounding clip...

Page 20: ...nted laterally 7 Commissioning 7 1 Preparatory steps 7 1 1 Factory settings The sensor is preconfigured on leaving the factory Once the installation parameters have been adjusted the sensor is ready f...

Page 21: ...ing When troubleshooting the entire measuring point must be taken into account Transmitter Electrical connections and cables Assembly Sensor The possible causes of error in the following table refer p...

Page 22: ...ety gloves Clean away splashes on clothes and other objects 10 1 Maintenance tasks 10 1 1 Cleaning the sensor Sensor without wiper Sensor fouling can affect the measurement results and even cause a ma...

Page 23: ...ft assembly Socket O ring Shaft Washer 71156833 For more detailed information on spare parts kits please refer to the Spare Part Finding Tool on the Internet www products endress com spareparts_consum...

Page 24: ...ax CUS71D 24 Endress Hauser 11 4 Disposal The device contains electronic components and must therefore be disposed of in accordance with regulations on the disposal of electronic waste Observe the loc...

Page 25: ...dular assembly system for sensors in open basins channels and tanks Material PVC or stainless steel Product Configurator on the product page www endress com cya112 Technical Information TI00432C PVC s...

Page 26: ...CUS71D ultrasonic sensor 12 2 Holder Flexdip CYH112 Modular holder system for sensors and assemblies in open basins channels and tanks For Flexdip CYA112 water and wastewater assemblies Can be affixed...

Page 27: ...r 71145499 13 Technical data 13 1 Input 13 1 1 Measured values Interface 13 1 2 Measuring ranges 0 3 to 10 m 1 to 32 ft 13 2 Performance characteristics 13 2 1 Reference operating conditions 20 C 68 F...

Page 28: ...4 1 Process temperature 1 to 50 C 34 to 122 F 13 4 2 Process pressure Standard version 0 to 6 bar 0 to 87 psi Version with wiper 0 to 0 3 bar 0 to 4 4 psi 13 5 Mechanical construction 13 5 1 Dimension...

Page 29: ...ming acceptance 8 Installation 11 IT security 7 M Maintenance 22 Measuring principle 8 Measuring system 14 Pendulum holder 16 N Nameplate 9 O Operational safety 7 P Post connection check 20 Product de...

Page 30: ......

Page 31: ......

Page 32: ...www addresses endress com 71415440 71415440...

Reviews: