Operation
Turbimax CUE21 / CUE22
26
Hauser
5. Press the key until OFST is displayed in the lower line.
At this point, the upper line of the display indicates the operating status of the
offset function. If the function is switched off, switch it on by pressing or .
6. The offset is the difference between the NTU value measured by the device and the
value measured by the lab device.
Example:
CUE21 / CUE22 measures 0.016 NTU,
while the lab device calculates a value of 0.012 NTU for the same sample. Entering
an offset of -0.04 results in the CUE21 / CUE22 also displaying 0.012 NTU.
Select the desired offset value using the or keys. Press the key to accept the
value.
This completes offset configuration. The device is still in configuration mode (CONFIG).
Press the MODE/EXIT key to return to AUTO mode.