6 Commissioning
Hauser
59
6.6.4
Set up 1 – Current outputs
The transmitter is always equipped with two current outputs.
To enter the menu, proceed as follows:
⇒
⇒
CHOICE
(default = bold)
INFO
Current output 1
Current output 2
Select a current output
for which the settings apply.
Current output 1 (or 2):
pH/Redox
Temperature
Controller (current output 2 only)
Selecting parameters
which should be output at the current output.
Selection possibilities related to the instrument variant and the
selected output.
Current output 1
(terminals 31+, 32-):
– pH/redox
– Temperature
Current output 2
(terminals 33+, 34-):
– pH/redox
– Temperature
– Controller: Control of a controlling actuator via the current
output (see also Controller menu page 74).
!
Note!
Danger of data loss. If you change the assignment for the current
output from "controller" to a different function
after
you have
configured the controllers, the
complete
controller settings
(see page 74) are reset to the default values.
Caution!
The configuration is changed.
Note in display (for changed setting):
Cancel by pressing "PARAM"
Continue (= confirm change) by pressing "E"
0 ... 20 mA
4 ... 20 mA
Current range selection
!
Note!
If you change the current range after editing the table, the table will
be reset.
!!Caution!!
Current output 0...20 mA and error current =
2.4 mA is dangerous.
Note in display:
Error current lies within the measuring current range. When the
current range is "0 ... 20 mA" and "Min" is selected under Alarm in
(see page 65).
Recommended combinations:
Current range 0...20 mA and error current max (22 mA)
or
Current range 4...20 mA and error current min (2.4 mA)
Linear
Table
Characteristic selection
Linear:
The characteristic is linear from the lower to the upper
value.
Table:
When you do not want the current output characteristic to
be linear, you can enter a customer-specific sequence of up to 10
value pairs in a table. Exact adaptation to the non-linear medium
behaviour can achieve a higher level of accuracy.
PARAM
pH
7.00
Param
Parametrieren
Grundeinstellung
Sonderfunktionen
Handbedienung
Quick
Setup
Edit
Weiter E
pH
7.00
Hold
Param
Settings
Set up 1
Set up 2
Manual operation
First start up
Edit
Next E
pH
7.00
Param
Set up 1
Sensor input
Display
Access codes
Current output
Relays
Edit
Next E
Hold
Hold
pH
7. 00
Param
Edit
[¯]
[E]
Output 2
pH/mV Input 1
Temperature Input1
Controller
Next
Summary of Contents for Topclean S CPC30
Page 154: ...10 Technical data Topclean S 154 Endress Hauser...
Page 156: ...11 Appendix Topclean S 156 Endress Hauser C07 CPC30xxx 19 06 08 en 006 eps...
Page 157: ...Topclean S 11 Appendix Endress Hauser 157 C07 CPC30xxx 19 06 08 en 007 eps...
Page 158: ...11 Appendix Topclean S 158 Endress Hauser C07 CPC30xxx 19 06 08 en 008 eps...
Page 159: ...Topclean S 11 Appendix Endress Hauser 159...
Page 160: ...11 Appendix Topclean S 160 Endress Hauser C07 CPC30xxx 19 06 08 en 002 eps...
Page 161: ...Topclean S 11 Appendix Endress Hauser 161 C07 CPC30xxx 19 06 08 en 003 eps...
Page 162: ...11 Appendix Topclean S 162 Endress Hauser C07 CPC30xxx 19 06 08 en 014 eps...
Page 163: ...Topclean S 11 Appendix Endress Hauser 163 C07 CPC30xxx 19 06 08 en 015 eps...
Page 164: ...11 Appendix Topclean S 164 Endress Hauser C07 CPC30xxx 19 06 08 en 011 eps...
Page 165: ...Topclean S 11 Appendix Endress Hauser 165 C07 CPC300xx 19 06 08 en 010 eps...
Page 166: ...11 Appendix Topclean S 166 Endress Hauser C07 CPC30xxx 19 06 08 en 012 eps...
Page 167: ...Topclean S 11 Appendix Endress Hauser 167 C07 CPC30xxx 19 06 08 en 013 eps...
Page 176: ...Index Topclean S 176 Endress Hauser...
Page 178: ...www endress com worldwide BA236C 07 en 11 04 51504339 Printed in Germany FM SGML 6 0 DT...