Analyzer, open version, SI-A and SI-B
a0001356
Open version (without housing)
With the open version, you need an additional platform for the reagents. Mount this platform max. 35 cm
(13.8 inch) below the pumps. The reagent bottels have the following dimensions: 90 x 90 x 215 mm (3.54 x
5.54 x 8.46 inch). The number of bottles varies from 2 to 5 depending on the analyzer version.
For these versions, the outlet pipe must be installed right of the analyzer.
The outlet pipe must be mounted to a wall so that the sample outlet hoses from the photometer have a gradient
of 5 to 10 %. If neccessary, extend the hoses.
RS 232
RS 232
381 (15.0)
351 (13.8)
325 (12.8)
8 (0.31)
9 (0.35)
540
(21.3)
562
(22.1)
19 (0.75)
197 (7.76)
192
(7.56)
175
(6.89)
mm (inch)
Ø6