StamoLys CA 71 SI
Installation
Hauser
13
3.4.2
CAT 411, CAT 430 and two CA 71 analyzers (two-channel
version)
3.5
Post-installation check
•
After installation, check that all connections are fitted tightly and are leakage resistant.
•
Ensure that the hoses can not be removed without effort.
•
Check all hoses for damage.
•
Permeate not free of air bubbles
•
Distance between the analyzers as short
as possible: sampling line between
T-piece and the second analyzer
(Fig. 11, item 5) shorter than 1.5 m
•
Cross-section of sampling line ID
3.2 - 4 mm
•
one sample receiver each
(without
level measurement)
for CAT 411 or
CAT 430
!
Note!
Ensure that there is always sufficient
sample available for both analyzers.
Observe this when selecting maintenance
intervals for CAT 411 and CAT 430.
C07-CA71xxx-11-08-00-xx-006.eps
Fig. 11:
Installation example
1
Sample from CAT 430
2
Sample from CAT 411
3
Collecting vessel
4
Collecting vessel overflow
5
Sampling lines
6
T-pieces
1
CA 71
CA 71
2
6
3
4
5