BA02195C
OXY5500 Optical Oxygen Analyzer
Hauser
4
–
41
1.
Change the values for
Cal0
,
T0
,
Cal2nd
,
T2nd
and
pATM
according
to the values shown in the Calibration Certificate. Refer to Figure
4–41.
2.
Change the
O2-2nd
value according to the value shown below the
cal2nd column in Figure 4–34.
3.
Click the Save button in the upper right of the screen to store the
changes and complete the manual calibration of the analyzer.
The display will automatically change to the MEASUREMENT window.
If another sensor type has been selected, a message window will
display stating that the sensor type change has reset RATA (refer to
“Relative accuracy test audit (RATA)”
46). Refer to
Performing a two-point calibration
To perform a two-point calibration with the connected oxygen sensor, begin by
selecting the screens below. When completed, continue with the procedure
detailed in
35.
On the Calibration Certificate, “pATM” is shown as “Atmospheric
Pressure” under
Calibration Specifications
during Cal0 and
Cal2nd. Refer to Figure 4–34 on page 4–32
Confirm that the correct unit for the
O2-2nd
and
pATM
values
are selected.
Figure 4–41
Calibration data screen - Changing
the pressure unit