Solitrend MMP20 (Option D)
Commissioning
Hauser
21
safety and reliability when monitoring and checking a known type of concrete. It is
advisable for the user to document the particular varieties of concrete that are measured.
This makes it easier to verify the values during subsequent control measurements.
EC-T measuring range
• Concrete with low cement content or special cement types: 15 dS/m
• Concrete with higher cement content or special cement types: 45 dS/m
The EC-T parameter can only be evaluated properly if the type of concrete is known.
7.9
General settings
Changing the settings:
1. Press the Folder key for longer (2 s)
"Settings" menu
2. Press the navigation keys
Navigate to a menu item
3. Press the Enter key
Confirm the selected menu item
4. Press the Folder key
Exit the current menu item and the "Settings" menu
7.9.1
Overview of the configuration options
•
Find probe
Searches for a connected probe
•
Language
Change the system language
• German
• English
•
Auto-power-off
Setting for automatic switch-off
•
Display lighting
Setting for background lighting
• Switch-off time
• Brightness
•
Display contrast
Setting for optimum contrast
•
Probe info
Displays information about the probe
•
Info
Displays information about the handheld device
•
Material calibration
Select calibration curve for different materials
7.9.2
Find probe
Select the "Find probe" menu item if:
• There are communication problems between the handheld device and probe during
power-up
• The probe is connected for the first time
• The probe is to be changed during operation
Once this menu item has been selected, the handheld device makes another attempt to
establish a connection to a connected probe.