Solitrend MMP20 (Option D)
Commissioning
Hauser
15
7
Commissioning
7.1
Checking the packaging contents
• Handheld device
• Power adapter (12 V/2 A)
• Charging adapter
• Protection cap
• Manual
• SWZ probe
7.2
Charging the battery
Charging the battery before using the probe for the first time
1. Insert the charging adapter into the 7-pin socket on the handheld device
2. Connect the power adapter to the charging adapter
Charging commences immediately if the device is already switched on or the
battery is over-discharged.
3. Otherwise, switch on the device by pressing the Enter key for approximately 1 s
An animated battery symbol on the display indicates that charging is active. The
built-in charging electronics charge the battery until it is fully charged. Charging
takes around 2 h if the battery was fully discharged. As soon as the battery is
finished charging, all 4 "battery bars" appear permanently on the display screen
and trickle charging commences.
Only charge the battery at room temperatures.
If the temperature is too low, end-
of-charge cutoff might not work correctly and the battery may be overcharged. If
ambient temperatures are too high, the device can be damaged by the heat generated
during charging.
7.3
Connecting the probe
1. Insert the probe into the 7-pin socket on the device
2. Tighten the coupling nut
7.4
Switching the handheld device on/off
1. Press the Enter key for approximately 1 s
The device tries to communicate with the connected probe during the power-up
process. This takes around 4 s. If no probe is connected or if the probe cannot be
found for some other reason, an error message appears on the screen. If the
device managed to successfully find the probe, the measurement background
appears on the screen, depending on the specific operating mode. The message
"Calibrating" appears on the bottom of the display during the power-up process.
The device adjusts to the probe.
2. The device is now ready for use
3. Switch off the device
Press the Enter key for approximately 1 s