Wiring
Smartec S CLD134
4
Hauser
2.2
Installation of the compact version or CLS54 sensor for separate
version
!
Note!
Perform an air set and calibrate the sensor before installing the compact version or the sensor.
Install the compact version or the CLS54 sensor directly on the pipe or vessel socket via the
process connection (depending on ordered version).
1.
When installing the Smartec S CLD134 or the sensor, make sure that the flow opening of
the sensor is oriented in the flow direction of the medium. An orientation arrow on the
sensor facilitates orientation.
2.
Tighten the flange.
3
Wiring
#
Warning!
• The electrical connection must only be carried out by a certified electrician.
• Technical personnel must have read and understood the instructions in this manual and must
adhere to them.
• Ensure that there is no voltage at the power cable before beginning the connection work.
3.1
Cable glands
a0005439
Fig. 2: Terminal assignment of cable glands
A
1
2
3
4
5
6
Separate version
Cable gland for analog output, binary input
Cable gland for alarm contact
Cable gland for power supply
Housing ground
Pressure comp. element PCE (Goretex
®
- filter)
Cable gland for sensor connection, M 16x1.5
B
1
2
3
4
5
Compact version
Cable gland for analog output, binary input
Cable gland for alarm contact
Cable gland for power supply
Housing ground
Pressure comp. element PCE (Goretex
®
- filter)
A
1
2
3
4
5
6
B
1
2
3
4
5