KA401C/07/en/07.07
71059831

valid from:
Software version 1.13

Brief Operating Instructions

Smartec S CLD134

Conductivity Measuring System

These instructions are Brief Operating Instructions.
For detailed information, please read the Operating Instructions.

The complete device documentation comprises:
• these Brief Operating Instructions
• the Operating Instructions

Summary of Contents for Smartec S CLD134

Page 1: ...structions Smartec S CLD134 Conductivity Measuring System These instructions are Brief Operating Instructions For detailed information please read the Operating Instructions The complete device docume...

Page 2: ...d media It is particularly suitable for use in the foodstuffs industry Any other use than the one described here compromises the safety of persons and the entire measuring system and is therefore not...

Page 3: ...te version Wall mounting For wall mounting attach the mounting plate to the wall by drilling holes as required Anchors and screws are to be provided by the operator Note Wall mounting is not recommend...

Page 4: ...ng Warning The electrical connection must only be carried out by a certified electrician Technical personnel must have read and understood the instructions in this manual and must adhere to them Ensur...

Page 5: ...tput 2 temperature Auxiliary power output Binary input 2 MRS 1 2 Binary input 1 hold MRS 3 4 F G H I MRS Conductivity sensor Temperature sensor Alarm contact position no current Power supply Remote pa...

Page 6: ...not energized S2 External contacts not energized 3 3 Structure and termination of measuring cable 85 S1 D1 15 V A B D2 86 81 82 93 94 S2 a0006037 en Fig 5 Structure of the measuting cable a0006036 Fi...

Page 7: ...ve on within function groups and In setup mode the PLUS and MINUS keys have the following functions Selection of function groups Setting of parameters and numeric values In measuring mode repeatedly p...

Page 8: ...simultaneously to quit the Service function group 7 Press the MINUS key several times until the Setup 1 function group is displayed 8 Press the ENTER key to edit Setup 1 SETUP 1 9 In A1 select the op...

Page 9: ...eal temperature 35 0 250 0 C 0 0 C RealTemp 22 The difference between the measured and the entered temperatures is displayed Press the ENTER key The display returns to the initial display of the Setup...

Page 10: ...not o k the second display line shows an explanation of the error C114 Store calibration results yes no new yes Store If C113 E xxx then only no or new If new return to C If yes no return to Measurem...

Page 11: ...mperature 25 C 35 0 250 0 C 25 0 C MTC temp Only exists if B1 fixed C132 Enter value of calibration solution 2 10 K 0 00 20 00 K 2 10 K alpha value This value is specified in the Technical Information...

Page 12: ...www endress com worldwide KA401C 07 en 07 07 Printed in Germany FM SGML 6 0 71059831...