Level calibration probe 1
Step
Matrix
Entry
Significance
1
V0H1
e.g. 0.5 m With level at minimum, enter level in %, m, ft etc
2
-
»E«
Register entry
3
V0H2
e.g. 3 m
With level at maximum, enter level in %, m, ft etc.
4
-
»E«
Register entry
Level calibration probe 2
Step
Matrix
Entry
Significance
1
V4H1
e.g. 2 m
With level at minimum, enter level in same units as used
for probe 1
2
-
»E«
Register entry
3
V4H2
e.g. 5 m
With level at maximum, enter level in same units as used
for probe 2
4
-
»E«
Register entry
5
V0H0
…
Indicates level measured by probe 1
V4H0
…
Indicates level difference h
2
- h
1
Note!
•
The »empty« and «full« calibrations can be carried out in any order.
•
Configure analogue outputs, see Chapter 6.
-
Output 1 measures level h
1
measured by probe 1
-
Output 2 measures differential level h
2
- h
1
•
Configure limit switches, see Chapter 7.
Silometer FMC 672 Z/677 Z
Chapter 8: Other Applications
43