Installation
Proline t-mass T 150
24
Hauser
6.2.5
Turning the display module
Esc
A0017228
6.3
Post-installation check
Is the device undamaged (visual inspection)?
Does the measuring device conform to the measuring point specifications?
For example:
• Process temperature
• Process pressure (refer to the chapter on "Material load curves" of the "Technical Information"
document on the CD-ROM provided)
• Ambient temperature (→ 20)
• Measuring range
Has the correct orientation for the sensor been selected (→ 13)?
• According to sensor type
• According to medium properties
• According to medium temperature
• According to process pressure
Does the arrow on the sensor match the direction of flow of the medium through the piping (→ 13)?
Have sufficient inlet and outlet runs been provided upstream and downstream of the measuring point?
Correctly aligned in the direction of flow?
Is the device adequately protected from precipitation and direct sunlight?
Is the device protected against overheating?
Is the device protected against excessive vibrations?
Check liquid properties (e.g. purity, cleanness).
Are the measuring point identification and labeling correct (visual inspection)?