Proline Prosonic Flow P 500
Hauser
47
Sensor installation – medium/large nominal diameters DN 50 to 4000 (2 to 160")
Installation for measurement via 1 traverse
Requirements
• The installation distance and wire length are known → 39
• Strapping bands are pre-assembled
Material
The following material is required for mounting:
• Two strapping bands incl. mounting bolts and centering plates where necessary (already pre-
• Two measuring wires, each with a cable lug and a fixer to fix the strapping bands
• Two sensor holders
• Coupling medium (coupling pad or coupling gel) for an acoustic connection between the sensor
and pipe
• Two sensors incl. connecting cables
Installation is unproblematic up to DN 400 (16"), as of DN 400 (16") check the distance and
angle (180°) diagonally with the wire length.
Procedure:
1. Prepare the two measuring wires: arrange the cable lugs and fixer such that the distance they
are apart corresponds to the wire length (SL). Screw the fixer onto the measuring wire.
1
2
3
4
5
6
7
8
9
10
11
SL
A0043379
37
Fixer and cable lugs at a distance that corresponds to the wire length (SL)
2. With measuring wire 1: fit the fixer over the mounting bolt of strapping band 1 that is already
securely mounted. Run measuring wire 1 clockwise around the measuring pipe. Fit the cable
lug over the mounting bolt of strapping band 2 that can still be moved.
3. With measuring wire 2: fit the cable lug over the mounting bolt of strapping band 1 that is
already securely mounted. Run measuring wire 2 counterclockwise around the measuring pipe.
Fit the fixer over the mounting bolt of strapping band 2 that can still be moved.
4. Take the still movable strapping band 2, incl. the mounting bolt, and move it until both
measuring wires are evenly tensioned and then tighten strapping band 2 so that it cannot slip.
Then check the sensor distance from the center of the strapping bands. If the distance is too
small, release strapping band 2 again and position it better. Both strapping bands should be as
perpendicular as possible to the measuring pipe axis and parallel to one another.
A0043380
38
Positioning the strapping bands (steps 2 to 4)
5. Loosen the screws of the fixers on the measuring wires and remove the measuring wires from
the mounting bolt.