Description of device functions
Proline Prosonic Flow 91
104
Hauser
10.16
Group CALIBRATION DATA
Functional description CALIBRATION DATA
CAL. FACTOR
Use this function to call up the calibration factor currently used.
Data indicated
:
5-digit floating-point number (typically 1.000)
ZERO POINT
Use this function to call up the zero point correction currently used.
Data indicated
:
5-digit floating-point number with sign (e.g. +0200.0)
ZEROPOINT STAT.
Use this function to call up or manually change the static zero point correction currently
being used.
User input
:
5-digit floating-point number, including unit and sign (e.g. +0010.0 ns)
CORR. FACTOR
Use this function to enter a correction factor at the client's site.
User input
:
5-digit floating-point number between 0.5 and 2.
Factory setting
:
1.000 (no correction)