Proline Promass P 300
Hauser
43
DN
1:1
1:10
1:20
1:50
1:100
1:500
[mm]
[kg/h]
[kg/h]
[kg/h]
[kg/h]
[kg/h]
[kg/h]
40
45
000
4
500
2
250
900
450
90
50
70
000
7
000
3
500
1
400
700
140
US units
DN
1:1
1:10
1:20
1:50
1:100
1:500
[inch]
[lb/min]
[lb/min]
[lb/min]
[lb/min]
[lb/min]
[lb/min]
³⁄₈
73.50
7.350
3.675
1.470
0.735
0.147
½
238.9
23.89
11.95
4.778
2.389
0.478
1
661.5
66.15
33.08
13.23
6.615
1.323
1½
1
654
165.4
82.70
33.08
16.54
3.308
2
2
573
257.3
128.7
51.46
25.73
5.146
Accuracy of outputs
The outputs have the following base accuracy specifications.
Current output
Accuracy
±5 µA
Pulse/frequency output
o.r. = of reading
Accuracy
Max. ±50 ppm o.r. (over the entire ambient temperature range)
Repeatability
o.r. = of reading; 1 g/cm
3
= 1 kg/l; T = medium temperature
Base repeatability
Mass flow and volume flow (liquids)
±0.05 % o.r.
Density (liquids)
±0.00025 g/cm
3
Temperature
±0.25 °C ± 0.0025 · T °C (±0.45 °F ± 0.0015 · (T–32) °F)
Response time
The response time depends on the configuration (damping).
Influence of ambient
temperature
Current output
Temperature coefficient
Max. 1 μA/°C
Pulse/frequency output
Temperature coefficient
No additional effect. Included in accuracy.