Diagnostics and troubleshooting
Proline Promass E 300 Modbus RS485
150
Hauser
Options
Description
Logbook entry only
The device continues to measure. The diagnostic message is displayed only in the
Event
logbook
submenu (
Event list
submenu) and is not displayed in alternation with the
operational display.
Off
The diagnostic event is ignored, and no diagnostic message is generated or entered.
12.8 Overview of diagnostic information
The amount of diagnostic information and the number of measured variables affected
increase if the measuring device has one or more application packages.
In the case of some items of diagnostic information, the diagnostic behavior can be
changed. Change the diagnostic information → 149
Diagnostic
number
Short text
Remedy instructions
Status
signal
[from the
factory]
Diagnostic
behavior
[from the
factory]
Diagnostic of sensor
022
Temperature sensor
defective
1. Check or replace sensor electronic
module (ISEM)
2. If available: Check connection cable
between sensor and transmitter
3. Replace sensor
F
Alarm
046
Sensor limit exceeded
1. Inspect sensor
2. Check process condition
S
Warning
062
Sensor connection faulty 1. Check or replace sensor electronic
module (ISEM)
2. If available: Check connection cable
between sensor and transmitter
3. Replace sensor
F
Alarm
063
Exciter current faulty
1. Check or replace sensor electronic
module (ISEM)
2. If available: Check connection cable
between sensor and transmitter
3. Replace sensor
S
Alarm
082
Data storage
1. Check module connections
2. Change electronic modules
F
Alarm
083
Memory content
1. Restart device
2. Restore HistoROM S-DAT backup
('Device reset' parameter)
3. Replace HistoROM S-DAT
F
Alarm
140
Sensor signal
asymmetrical
1. Check or replace sensor electronic
module (ISEM)
2. If available: Check connection cable
between sensor and transmitter
3. Replace sensor
S
Alarm
144
Measurement error too
high
1. Check or change sensor
2. Check process conditions
F
Alarm
Diagnostic of electronic
201
Device failure
Restart device
F
Alarm
242
Software incompatible
1. Check software
2. Flash or change main electronics
module
F
Alarm