15
Excample for max. measured error
a0003401
Max. measured error in % o.r. (example: Promass 80A, 83A / DN 2)
Flow values (example)
Design fundamentals
Repeatability
Design fundamentals
o.r. = of reading
Mass flow and volume flow (liquids)
Promass 83A, 80A: ±0.05% o.r.
Mass flow (gases)
Promass 83A, 80A: ±0.25% o.r.
Density (liquids)
±0.00025 g/cc
1 g/cc = 1 kg/l
Temperature
±0.25 °C ± 0.0025 · T °C
(±1 °F ± 0.003 · (T – 32) °F)
T = medium temperature
0
±0.5
±0.1
[%]
0
50
100
kg/h
±0.2
±0.3
±0.4
10
5
20
30
40
60
70
80
90
Turn down
Flow
Max. measured error
[kg/h]
[lb/min.]
[% o.r.]
250:1
0,4
0,0147
1,250
100:1
1,0
0,0368
0,500
25:1
4,0
0,1470
0,125
10:1
10
0,3675
0,100
2:1
50
1,8375
0,100
o.r. = of reading